IR Thermometer with Wireless USB Interface 

Product Code: MIT-371 

 

User Manual 

 

 

 

 

WWW.MARMONIX.CO 

 

 

Содержание MIT-371

Страница 1: ...IR Thermometer with Wireless USB Interface Product Code MIT 371 User Manual WWW MARMONIX CO...

Страница 2: ...Buttons 10 Measurement Operation 11 Data Logger 11 1 Storing Data 11 2 Infrared 11 3 Recalling Data 11 4 Log Clear Function 11 5 Transmit The Data Logger 11 6 Wireless Transmit Function 12 Battery Rep...

Страница 3: ...contact temperature measurement Type K temperature Measurement Dual laser sighting 150ms faster sampling time Wireless USB Interface RF 433Hz Unique flat surface modern housing design Automatic Data...

Страница 4: ...below Te focal point for each unit is 914mm 36 The spot sizes indicate 90 encircled energy CAUTION LASER RADIATION AVOID DIRECT EYE EXPOSURE MAXIMUM OUTPU 1Mw WAVELENGTH 630 670nm CLASS LASER PRODUCT...

Страница 5: ...ponse 8 14um Emissivity Digitally adjustable from 0 1 to 1 0 Over range Indication LCD will show OL OL Polarity Automatic no indication for positive polarity Minus sign for negative polarity Diode Las...

Страница 6: ...Auto obtain Emissivity 6 Lock and laser on symbols 7 High alarm and low alarm symbol 8 Temperature values for the MAX MIN DIF AVG HAL LAL LAL TK and LOG 9 Symbols for EMS MAX MIN DIF AVG HAL LAL TK An...

Страница 7: ...witching C F Lock ON OFF and Set Alarm 1 C F 2 LOCK ON OFF 3 Set Alarm Select the temperature units C or F by using C Fswitch To lock the unit on for continuous measurement slide the middle switch LOC...

Страница 8: ...e LCD IR temperature value will be in the middle of LCD and Type K temperature value is at the lower side of the LCD Contact the Type K probe to the object surface and test the temperature of same poi...

Страница 9: ...rs in the lower left corner A LOG location number will be shown below LOG and the stored temperature for that location will be display To move to another LOG location press the UP and Down keys 11 4 L...

Страница 10: ...urned off After the Wireless Transmit function is turned on firstly connect the USB interface RF 433MHz to the computer and wireless transmitter and open the software on the computer then the IR tempe...

Страница 11: ...ture of an object The unit s optics sense emitted reflected and transmitted energy which is collected and focused on to a detector The unit s electronics translate the information into a temperature r...

Страница 12: ...s It will measure the surface temperature of the glass instead Steam dust smoke etc can prevent accurate measurement by obstructing the unit s optics 14 3 Emissivity Emissivity is a team used to descr...

Страница 13: ...94 Snow 0 83 Plastic 0 85 to 0 95 Glass 0 90 to 0 95 Timber 0 90 Ceramic 0 90 to 0 94 Paper 0 70 to 0 94 Marble 0 94 Chromium oxides 0 81 Plaster 0 80 to 0 90 Copper oxides 0 78 Mortar 0 89 to 0 91 Ir...