Trouble Shooting
1. The Meter is displaying the wrong readings
Enter the calibration mode and correct using a known trusted source
such as a reputable internet source for Barometric and Altitude
Readings. Or a known accurate stable air flow will need passing
through the meter vane in working mode (an accredited wind flow
tunnel is strongly recommended), to assist in calibrating the wind
speed.
2. Power on but no display or the display disappears
a. Check the battery is in place with good contact to the pins and
the correct way up.
b. Remove battery for one minute and re-install.
c. Replace the battery.
3. For all other issues please contact your supplier.
Relative Altitude Measurement
To view the relative altitude, or the change in altitude since
last calibration (referred to as climb or ascent / descent),
press and hold the Units Button for two seconds, the LCD
will then display the REL interface on screen and then rela-
tive altitude (Defaulted value 0m). The value on the screen
will change as the Anemometer is moved higher or lower
than its current position. Press and hold the Units Button
again for two seconds to change back to Absolute (Density)
Altitude. Please see below picture:
Battery Replacement
If the meter does not power on as usual or the low battery
icon
appears on the LCD display, please replace the
batteries with a new 9V battery as soon as possible.
Dispose of the battery in an environmentally safe at a recy-
cling point.
DO NOT DISPOSE IN YOUR DUST BIN!
Calibration function
Step 1.Hold Units Button then turn on meter to enter cali-
bration mode.
Step 2: Press Units button choose the mode “see table
below”, then press Power Button to enter the calibration
mode;
Step 3: Press Units Button to choose the unit you wish to
change (the selected unit will flash quickly) and press Set
Button to adjust to the correct value,
Step 4: Press Power Button for over 2 seconds to store the
setting and exit.
Number
Calibration Function Mode
1
Wind Speed
2
Temperature
3
Relative Humidity
4
Barometric Pressure
5
Absolute (Density) Altitude
Caution:
1. This meter is accurate at low and mid-range air speeds. Consistent
use at very high speeds, i.e. from the window of a fast moving car
will damage the impeller’s bearing and reduce accuracy.
2. This meter is not a personal safety device.
Warranty
This meter has a one year warranty for a period of one year from
purchase date.
This warranty covers normal operation and does not cover batteries
(including leaks), misuse, abuse, alteration, tampering, neglect, im-
proper maintenance or improper calibration. Nor can the warranty be
used as an excuse for a false claim simply because you have changed
your mind after the initial 14 day period.
Proof of purchase is required for warranty repairs.