Introduction 

1.1  Notices 

This manual was prepared exclusively for the owner of the Markes International ACTI-VOC Low-flow pump. The 
material within this manual is proprietary information and is to be used only to understand, operate, and 
maintain the instrument. By receiving this document, the recipient agrees that neither this document nor the 
information disclosed within nor any part thereof shall be reproduced or transferred, physically, electronically or 
in any other form, or used or disclosed to others for manufacturing or for any other purpose except as 
specifically authorised in writing by the manufacturer. 

1.2  Disclaimer 

The seller assumes no responsibility whatsoever, to any party whosoever, for any property damage, personal 
injury, or death received by or resulting from, in whole, or in part, the improper use, installation, or storage of this 
product by the user, person, firm, entity, corporation or party not adhering to the instructions and warnings in 
this manual, or otherwise provided by the seller or from not adhering to all federal, state, and local 
environmental and occupational health and safety laws and regulations. 

The seller shall not be liable for direct, indirect, consequential, incidental or other damages resulting from the 
sale and use of any goods and sellers’ liability hereunder shall be limited to repair or replacement of any goods 
found defective. This warranty is in lieu of all other warranties, expressed or implied, including but not limited to 
the implied warranties of merchantability and fitness for use or for a particular purpose which are expressly 
disclaimed. 

1.3  Warnings 

READ AND UNDERSTAND ALL WARNINGS AND INSTRUCTIONS BEFORE USE 

 

Failure to read, understand, and comply with ALL accompanying literature, product labels, and warnings 
could result in property damage, severe personal injury, or death. 

 

Read and understand ALL applicable environmental health and safety laws and regulations. Ensure 
complete compliance with ALL applicable laws and regulations before and during use of this product. 

 

DO NOT

 remove, cover, or alter any label or tag on this product, its accessories, or related products. 

 

UNDER NO CIRCUMSTANCES

 should this product be used except by qualified, trained, technically 

competent personnel. 

 

DO NOT

 operate this product should it malfunction, require repair, or have a cracked or broken case. 

DO NOT

 repair or modify this product, except as specified in this manual. No user-serviceable parts are 

located within the case. Contact Markes International for service-related queries. 

 

Use only specified Markes parts when performing maintenance procedures described in this manual. 
Intrinsic safety certifications become void by substitution of components, unauthorised repair or 
alteration. 

 

This product is intended for both indoor and outdoor use when protected from splashed or wind-blown 
liquids. The unit is not waterproof, so 

NEVER

 submerge the unit in water. Pump failure or faulting may 

result. 

 

This product presents a possible static hazard. Do not rub the product with a dry cloth. 

 

DO NOT

 charge the battery in an explosive atmosphere. 

 

This product uses rechargeable nickel-metal-hydride batteries, which should always be fully charged 
before use. Use only the battery pack and chargers specified in this manual. 

DO NOT

 insert any foreign 

objects into the contact slot. Shorting contacts will blow the protective fuse. 

 

If the equipment is likely to come into contact with aggressive substances, then it is the responsibility of 
the user to take suitable precautions that prevent it from being adversely affected. Examples of 

Содержание ACTI-VOC

Страница 1: ...Markes International Ltd T 44 0 1443 230935 F 44 0 1443 231531 E enquiries markes com www markes com ACTI VOC Low flow pump Operator Manual Version 1 0 QUI 1086 March 2013 ...

Страница 2: ...contact details 4 2 Components 5 3 Pump operation 7 3 1 Setting a constant flow rate 7 3 2 Using constant pressure mode 9 3 3 Taking a personal sample 10 4 Maintenance 11 4 1 Battery maintenance 11 4 2 Filter maintenance 11 4 3 Back pressure adapter O ring 12 5 Parts list 13 5 1 Accessories 13 5 2 Spare parts 13 6 Specifications 14 ...

Страница 3: ... Failure to read understand and comply with ALL accompanying literature product labels and warnings could result in property damage severe personal injury or death Read and understand ALL applicable environmental health and safety laws and regulations Ensure complete compliance with ALL applicable laws and regulations before and during use of this product DO NOT remove cover or alter any label or ...

Страница 4: ...o resolve your query please contact Markes International details below Address Gwaun Elai Medi Science Campus Llantrisant RCT CF72 8XL UK Website www markes com E Mail techsupport markes com Tel 44 0 1443 230935 Fax 44 0 1443 231531 Explosion protection equipment in compliance with Ex IEC 60079 11 AEx FM 3610 ib Intrinsic safety protection method IIC Gas group T4 135 C maximum external surface tem...

Страница 5: ...e screw cap prevents dirt from entering the Outlet port when not in use 9 Bag sampling accessory When installed into the Outlet port this accessory provides a means for filling air sampling bags 10 Case screw These screws must remain tight and should be checked prior to beginning sampling 11 Charging jack The Charging jack receptacle is used to connect a charger for recharging the internal battery...

Страница 6: ...6 www markes com Figure 1 Front view of ACTI VOC Figure 2 Back view of ACTI VOC ...

Страница 7: ...Continue turning the key until the mode indicator shows black for constant flow Turn the key clockwise to lock your selection Figure 3 Setting the pump to constant flow mode Attach tubing to the pump inlet Connect a TD tube to the tubing via the appropriate adapter Use the back pressure adapter when setting flows up to 100 mL min use the plastic tube adapter assembly when setting flows in excess o...

Страница 8: ...d instrument such as a ballpoint pen Make sure the flow measuring device is on and working Use a flat bladed screwdriver to make flow adjustments on the pump itself When the desired flow rate has been reached turn off the pump and flow measuring device The pump is now ready for sampling ...

Страница 9: ... constant pressure Turn the key clockwise to lock your selection Attach tubing to the pump inlet Connect the TD tube to the tubing via the in line flow controller When setting a sampling flow with the pump in constant pressure mode the actual tube which is to be used for sampling should be used to set the flow Connect the TD tube to a suitable flow measuring device e g digital flowmeter or primary...

Страница 10: ...rrect direction towards the pump The tube can be held in position with a penclip Use a pointed instrument such as a ballpoint pen to turn on the pump When sampling is completed use a pointed instrument such as a ballpoint pen to turn off the pump Record the sample data Re cap the sample tube finger tight plus a quarter turn with the CapLok tool When the pump is turned off after a sample run the to...

Страница 11: ... before use NiMH batteries stored for extended periods should be recharged every 1 2 months to avoid complete discharge The battery pack has an estimated life of 300 500 charge discharge cycles depending on use The table below shows the estimated battery life based on usage level Pump usage Estimated battery life High 40 60 hours per week 1 1 years Medium 20 39 hours per week 1 2 years Low 20 hour...

Страница 12: ... implements as described below Use Markes O ring extraction tool or similar pointed implement to hook out the O ring taking care not to scratch the groove in which the O ring is seated nor to damage the white PTFE insert behind Figure 6 Back pressure adapter with O ring extraction tool Take the replacement O ring and push it into back pressure adapter by hand Use Markes O ring insertion tool or si...

Страница 13: ...ard low flow calibrator 230 V 5 2 Spare parts Part number Description SERZ 1076 ACTI VOC filter assembly SERZ 1070 ACTI VOC bag sampling accessory SERZ 0351 O ring extraction tool SERZ 0285 O ring insertion tool SERFLP 5002 ACTI VOC charger SERZ 1097 Plastic tube adapter assembly SERLFP 5003 Back pressure adapter C CPLOK CapLok tool U COV45 O Ring low emission 5 92 mm i d pk 10 ...

Страница 14: ...re up to 25 H2O Flow control 5 of set point or 3 mL min whichever is greater Battery type Rechargeable nickel metal hydride battery pack Battery output 4 8 V 720 mA h Charging Internal external with adapter Operating temperature 20 C to 45 C 4 F to 113 F Storage temperature 40 C to 45 C 40 F to 113 F Charging temperature 0 C to 45 C 32 F to 113 F Additional features Flow fault indication LED Batte...