Mark-10 MTT03-100E Скачать руководство пользователя страница 7

Series TT03 Digital Torque Gauges 

 

User’s Guide 

 
 

 

 

 

6

 

High / low limit 
indicators 

Correspond to the programmed set points. Indicator definitions are as follows: 

 

– the displayed value is greater than the upper load limit  

 

– the displayed value is between the load limits 

 

– the displayed value is less than the lower load limit 

Set points 

The programmed load limit values. Typically used for pass/fail type testing. 
One, two, or no indicators may be present, depending on the configuration 
shown in the 

Set Points

 menu item. 

 

 

4.2 Controls 

Primary 

Label 

Primary Function 

Secondary 

Label 

Secondary Function 

 

 

Powers the gauge on and off. Press 
briefly to power on, press and hold 
to power off. Active only when the 
home screen is displayed. 

ENTER 

Various uses, as described in the 
following sections. 

ZERO  

Zeroes the primary reading and 
peaks. 

  

(UP) 

Navigates up through the menu and 
sub-menus . 

MENU  

Enters the main menu. 

ESCAPE 

Reverts one step backwards through 
the menu hierarchy.  

MODE  

Toggles between measurement 
modes.  

 

(DOWN) 

Navigates down through the menu 
and sub-menus. 

DATA  

Transmits the current reading to an 
external device via the USB port. 

DIRECTION 

Toggles between clockwise and 
counter-clockwise directions while 
configuring set points and other menu 
functions. 

 
Note: 

Measurement units are configured through the menu. Refer to 

Changing The Units 

section for 

details. 
 

4.3 Menu navigation basics 

Most of the gauge

’s various functions and parameters are configured through the main menu. To access 

the menu press 

MENU

. Use the 

UP

 and 

DOWN

 keys to scroll through the items. The current selection is 

denoted with clear text over a dark background. Press 

ENTER

 to select a menu item, then use 

UP

 and 

DOWN

 again to scroll through the sub-menus. Press 

ENTER

 again to select the sub-menu item.  

 
For parameters that may be either selected or deselected, press 

ENTER

 to toggle between selecting and 

deselecting. An asterisk (

*

) to the left of the parameter label is used to indicate when the parameter has 

been selected. 
 
For parameters requiring the input of a numerical value, use the 

UP

 and 

DOWN

 keys to increment or 

decrement the value. Press and hold either key to auto-increment at a gradually increasing rate. When 
the desired value has been reached, press 

ENTER

 to save the change and revert back to the sub-menu 

item, or press 

ESCAPE

 to revert back to the sub-menu item without saving. Press 

ESCAPE

 to revert one 

step back in the menu hierarchy until back into normal operating mode. 
 
Refer to the following sections for details about setting up particular functions and parameters. 
 
 

Содержание MTT03-100E

Страница 1: ...Series TT03 DIGITAL TORQUE GAUGES User s Guide...

Страница 2: ...and operation instructions Dimensions and specifications are also provided For additional information or answers to your questions please do not hesitate to contact us Our technical support and engin...

Страница 3: ...ures should be performed before and during operation 1 Never operate the gauge if there is any visible damage to the AC adapter or the gauge itself 2 Ensure that the gauge is kept away from water or a...

Страница 4: ...cess 1 When battery life is greater than 75 the following indicator is present 2 When battery life is between 50 and 75 the following indicator is present 3 When battery life is between 25 and 50 the...

Страница 5: ...our thumb screws fastened into the appropriate holes in the rear half of the housing The torque sensor portion of the gauge shown in Figure 3 1 can be mounted directly to a Mark 10 torque test stand o...

Страница 6: ...the left indicates counter clockwise load If set points are enabled triangular markers are displayed for visual convenience This indicator reflects the actual load which may not correspond to the prim...

Страница 7: ...kwise directions while configuring set points and other menu functions Note Measurement units are configured through the menu Refer to Changing The Units section for details 4 3 Menu navigation basics...

Страница 8: ...ZERO will reset the value 5 3 Peak Counter clockwise PCCW Same as above but for counter clockwise readings 6 CHANGING THE UNITS The TT03 can display three different measurement units depending on the...

Страница 9: ...screen appears as follows Either one two or none of the set points may be enabled To toggle between the clockwise and counter clockwise directions press the DIRECTION key If two set points have been...

Страница 10: ...he value only Polarity same as above Invert Polarity Clockwise values have negative polarity counter clockwise values have positive polarity May be selected in addition to the Numeric Units Numeric On...

Страница 11: ...of digits after the decimal point is dependent of the gauge s capacity and resolution Polarity may be inverted or omitted as shown in the table above Any detected errors are reported back by means of...

Страница 12: ...ould be calibrated at 10 20 30 40 and 50 lbFin loads in each direction 2 To escape the Calibration menu at any time press ESCAPE The display appears as follows Selecting CANCEL will revert back to the...

Страница 13: ...nsor several times at full scale if possible then press ENTER 7 The display appears as follows Apply torque equal to the full scale of the instrument then press ENTER 8 After displaying PLEASE WAIT th...

Страница 14: ...been completed the display appears as follows Press ENTER 11 At the completion of the counter clockwise calibration the display appears as follows To save the calibration information select SAVE EXIT...

Страница 15: ...he display appears as follows Select Disabled to disable automatic shutoff Select Enabled to enable it The length of time of inactivity is programmed in minutes via the Set Minutes parameter Available...

Страница 16: ...LCD Contrast The contrast of the display may be adjusted Select LCD Contrast from the menu The screen appears as follows Press ENTER to modify the contrast Select a value from 0 to 25 25 producing th...

Страница 17: ...ay be found in the Specifications section The screen appears as follows 11 7 Information Welcome Screen The following screen is displayed at power up and can be accessed at any time by selecting Infor...

Страница 18: ...24 hours of continuous use Measurement units lbFin ozFin kgFmm Ncm depending on model USB output Individual data point output by pressing DATA button Configurable up to 115 200 baud Safe overload 150...

Страница 19: ...isabled Baud Rate 9 600 Data Format Numeric units Automatic Shutoff Enabled Minutes 5 Tones Keys Enabled Alerts Enabled Set Points Momentary Initial Mode Real Time Units Depends on gauge model 12 3 Ca...

Страница 20: ...ition to our standard line of products we can provide modifications and custom designs for OEM applications Our engineering team is eager to satisfy any special requirements Please contact us for furt...

Отзывы: