Mark-10 EK3-100 Скачать руководство пользователя страница 7

Series 3 Digital Force Gauges 

 

User’s Guide

 

 
 

 

 

 

6

 

       kgF 

 Kilogram-force 

       gF 

 Gram-force 

       N 

 Newton 

Note:

 not all gauge capacities measure in all the above units. Refer to the 

capacity / resolution table in the 

Specifications

 section for details. 

Mode 

The current measurement mode. Abbreviations are as follows: 
       RT 

 Real Time  

       PC 

 Peak Compression  

       PT 

 Peak Tension  

See 

Operating Modes

 section for details about each of these modes 

Battery / AC 
adapter indicator 

Either the AC adapter icon or battery power icon will be shown, depending on 
power conditions. Refer to the 

Power

 section for details.  

High / low limit 
indicators 

Correspond to the programmed set points. Indicator definitions are as follows: 

 

 the displayed value is greater than the upper force limit  

 

 the displayed value is between the limits 

 

 the displayed value is less than the lower force limit 

Set points 

The programmed force limits. Typically used for pass/fail type testing. 1, 2, or 
no signals may be present, depending on the configuration shown in the 

Set 

Points

 menu item. 

 

4.2 Controls 

Primary 

Label 

Primary Function 

Secondary 

Label 

Secondary Function 

 

 

Powers the gauge on and off. Press 
briefly to power on, press and hold 
to power off. Active only when the 
home screen is displayed. 

ENTER 

Various uses, as described in the 
following sections. 

ZERO  

Zeroes the primary reading and 
peaks. 

  

(UP) 

Navigates up through the menu and 
sub-menus . 

MENU  

Enters the main menu. 

ESCAPE 

Reverts one step backwards through 
the menu hierarchy.  

MODE  

Toggles between measurement 
modes.  

 

(DOWN) 

Navigates down through the menu 
and sub-menus. 

DATA  

Transmits the current reading to an 
external device, via the USB port. 

DIRECTION 

Reverses the display during 
calibration, and toggles between 
tension and compression directions 
while configuring set points and other 
menu items. 

 
Note: 

Measurement units are configured through the menu. Refer to the 

Changing The Units

 section for 

details. 
 

4.3 Menu navigation basics 

Most of the gauge’s various 

functions and parameters are configured through the main menu. To access 

the menu press 

MENU

. Use the 

UP

 and 

DOWN

 keys to scroll through the items. The current selection is 

denoted with clear text over a dark background. Press 

ENTER

 to select a menu item, then use 

UP

 and 

DOWN

 again to scroll through the sub-menus. Press 

ENTER

 again to select the sub-menu item.  

 
For parameters that may be either selected or deselected, press 

ENTER

 to toggle between selecting and 

deselecting. An asterisk (

*

) to the left of the parameter label is used to indicate when the parameter has 

been selected. 
 
For parameters requiring the input of a numerical value, use the 

UP

 and 

DOWN

 keys to increment or 

decrement the value. Press and hold either key to auto-increment at a gradually increasing rate. When 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание EK3-100

Страница 1: ...Series 3 DIGITAL FORCE GAUGES User s Guide www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...y and industrial environments This User s Guide provides setup safety and operation instructions Dimensions and specifications are also provided For additional information or answers to your questions...

Страница 3: ...s that should not be used with the gauge include potentially flammable substances or products items that can shatter in an unsafe manner and any other components that can present an exceedingly hazard...

Страница 4: ...n icon appears in the lower left corner of the display as follows If the AC adapter is not plugged in battery power drainage is denoted in a five step process 1 When battery life is greater than 75 th...

Страница 5: ...he housing An additional two holes are supplied for metric screws These holes are designed to accommodate screws in order to hold the gauge in place Mark 10 test stands include a set of thumb screws f...

Страница 6: ...sing ZERO or by powering the gauge off and on 3 Primary reading The current displayed force reading See Operating Modes section for details 4 Load bar Analog indicator to help identify when an overloa...

Страница 7: ...primary reading and peaks UP Navigates up through the menu and sub menus MENU Enters the main menu ESCAPE Reverts one step backwards through the menu hierarchy MODE Toggles between measurement modes D...

Страница 8: ...peak compression reading observed If the actual force decreases from the peak value the peak will still be retained in the primary reading area of the display Pressing ZERO will reset the value 5 3 Pe...

Страница 9: ...mpared to these limits 8 2 Configuration To configure set points select Set Points from the menu The screen appears as follows Either one two or none of the set points may be enabled To toggle between...

Страница 10: ...ettings select USB Settings from the menu The screen appears as follows Communication settings are permanently set to the following Data Bits 8 Stop Bits 1 Parity None Other settings are configured as...

Страница 11: ...r where the Line Feed 0x0A is ignored 9 3 Command Responses In response to the reading request command the gauge will return a string with the load data followed by a space then the load unit if enabl...

Страница 12: ...o escape the Calibration menu at any time press ESCAPE The display appears as follows Selecting Cancel will revert back to the Calibration setup Selecting Exit w o saving will return to the menu witho...

Страница 13: ...y appears as follows Apply a weight equal to the full scale of the instrument then press ENTER 9 After displaying Please wait the display appears as follows CALIBRATION COMPRESSION Gain adjust Apply f...

Страница 14: ...n points have been completed the display appears as follows Press ENTER 12 The display appears as follows Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees Press DIRECTI...

Страница 15: ...illating or vibrating in any manner Then try again The calibration weight does not match the set value The entered calibration point is too close to the previous point CALIBRATION TENSION Load too clo...

Страница 16: ...m the menu The display appears as follows Selection Description Off Backlight to be off upon powering on the gauge On Backlight to be on upon powering on the gauge Auto Backlight to be on upon powerin...

Страница 17: ...The screen appears as follows 11 5 Initial Mode This section is used to configure the initial mode upon powering on the gauge To access this parameter select Initial Mode from the menu The screen appe...

Страница 18: ...f up to 24 hours of continuous use Measurement units lbF gF kgF N depending on model Safe overload 200 of full scale display shows OVER at 110 and above Weight gauge only M3 012 M3 100 0 7 lb 0 33 kg...

Страница 19: ...olution Load Cell Deflection Model lbF kgF gF N Load Cell Deflection in mm M3 012 0 12 x 0 0001 50 x 0 05 0 5 x 0 0005 0 005 0 13 M3 025 0 25 x 0 0002 100 x 0 1 1 x 0 001 0 010 0 25 M3 05 0 5 x 0 0005...

Страница 20: ...Force Gauges User s Guide 19 12 4 Dimensions IN MM Thread Flat M3 012 M3 100 10 32 UNF 5 16 7 94 M3 200 M3 500 5 16 18 UNC 5 16 7 94 www GlobalTestSupply com Find Quality Products Online at sales Glob...

Отзывы: