37

 |

 | 

MAQUET Medical Systems USA 

ENDOSCOPIC VESSEL HARVESTING

 |

VASOVIEW 7 | Training Manual | English | v1 

Copyright MAQUET Gmbh

VASOVIEW 7 xB Endoscopic Vessel Harvesting System  

caution:

 Federal (USA) law restricts this device for sale by, or on the order of, a physician.

important:

 These Instructions for Use are designed to assist in the use of the VASOVIEW 7 xB 

Endoscopic Vessel Harvesting System. They are not a reference to endoscopic surgery or tech-
niques. 

indicaTiOnS:

 The VASOVIEW 7 xB Endoscopic Vessel Harvesting System is indicated for use in 

minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for pa-
tients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and con-
trolling bleeding through coagulation, and for patients requiring blunt dissection of tissue including 
dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and 
other structures in the extraperitoneal or subcutaneous extremity and thoracic space. Extremity 
procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coro-
nary artery bypass grafting and peripheral artery bypass or the radial artery for use in coronary 
artery bypass grafting. Thorascopic procedures include exposure and dissection of structures 
external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall.

cOnTraindicaTiOnS:

 The VASOVIEW 7 xB Endoscopic Vessel Harvesting System is contraindi-

cated in situations where minimally invasive surgery is contraindicated.

WarningS and prEcauTiOnS:

 1) Read all instructions carefully. Failure to properly follow the 

instructions, warnings and precautions may lead to serious surgical consequences or serious injury 
to the patient. 2) Minimally invasive surgical procedures should be performed only by individuals 
adequately trained and familiar with such surgical techniques. Consult medical literature regarding 
techniques, complications, and hazards prior to performance of these procedures. 3) Sterility: The 
VASOVIEW 7 xB system is sterile unless the package is damaged or opened. The method of steri-
lization is gamma irradiation. The VASOVIEW 7 xB system is designed for single use. Do not reuse 
or resterilize. 4) Before endoscopic instruments and accessories from different manufacturers are 
employed in a procedure, verify compatibility and ensure that electrical isolation and grounding of 
these instruments is not compromised. 5) A thorough understanding of the principles and tech-
niques involved in electrosurgical procedures is essential to avoid shock and burn hazards to both 
the patient and operator(s) and damage to medical instrumentation. 6) The VASOVIEW 7 xB Endo-
scopic Vessel Harvesting System is for use with the bipolar outputs of electrosurgical generators 
only! 7) FOR INTERMITTENT OPERATION ONLY: Do not apply continuous cautery energy. 8) Do not 
exceed 30 watts in any mode. 9) Use generators listed in the specified Mode and Setting Range 
only. 10) Do not use with Codman Malis generators. 11) Handle the Endoscope carefully to avoid 
breakage. Regularly check the orientation of the camera before advancing. 12) To avoid damage to 
delicate tissue, advance the cannula gently. 13) Always advance the C-Ring and BiSECTOR™ tool 
under endoscopic visualization. Ensure adequate visualization of BiSECTOR ends and surgical site 
prior to application of electrosurgical energy. 14) Always inspect the surgical site for hemostasis. 
If hemostasis is not present, appropriate techniques should be applied to achieve hemostasis. 15) 
In endoscopic procedures which use gas insufflation, venous gas embolism is a very rare (approx-
imately 1 in 10,000 cases) but potentially serious complication that may occur. Its occurrence is 
signaled by cardiovascular collapse (sudden, severe hypotension) and a precordial murmur. If gas 
embolism is suspected during a procedure, discontinue gas insufflation and place the patient in a 
left lateral and a slight Trendelenburg position. 16) When performing radial artery harvesting, the 
radial artery harvesting procedure should be performed prior to placing the patient on cardiopul-
monary bypass. 17) Do not touch the coagulating surfaces while the device is activated. This may 
cause injury.

VASOVIEW Short Port Blunt Tip Trocar (BTT)

caution:

 Federal (USA) law restricts this device to sale by, or on the order of, a physician.

important:

 These Instructions for Use are designed to assist in the use of the Short Port Blunt Tip 

Trocar (BTT). They are not a reference to endoscopic surgery or techniques.  

indicaTiOnS:

 The VASOVIEW Short Port BTT has applications for surgery in the saphenous vein 

or radial artery for establishment of a port of entry for endoscopic instruments.

cOnTraindicaTiOnS:

 The VASOVIEW Short Port BTT is contraindicated in situations where mini-

mally invasive surgery is contraindicated.

WarningS and prEcauTiOnS:

 1) Sterility: The product is sterile unless the package is dam-

aged or opened. The method of sterilization is gamma irradiation. The product is designed for sin-
gle use. Do not reuse or resterilize. 2) Balloon products must be treated with care. Damage to the 
balloon by instruments used during insertion and in the course of a procedure may result in balloon 
rupture. 3) The Short Port BTT balloon contains natural rubber latex that is encapsulated by a sili-
cone coating such that latex is not intended to come in contact with the patient or the user. If the 
outer layer of the balloon on the Short Port BTT is damaged, natural rubber latex may be exposed. 
Natural rubber latex may cause allergic reactions. 4) Overinflation of the Short Port BTT balloon 
may result in balloon rupture. Do not inflate with more than 25 cc of air.

7 mm Extended Length Endoscope and Dissection Tip  

caution:

 Federal (USA) law restricts this device for sale by, or on the order of, a physician. 

indicaTiOnS:

 The VASOVIEW 7 mm Extended Length Endoscope with Dissection Tip is indicated 

for visualization of a surgical cavity and dissection in endoscopic procedures and other minimally 
invasive surgical procedures allowing access for vessel harvesting, and is primarily indicated for 
patients undergoing endoscopic vessel harvesting for arterial bypass. It is indicated for patients 
requiring endoscopic tissue separation of the extraperitoneal or subcutaneous extremity and tho-
racic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous 
vein for use in coronary artery bypass grafting and peripheral artery bypass or radial artery for use 
in coronary artery bypass.

cOnTraindicaTiOnS:

 The 7 mm Extended Length Endoscope and Dissection Tip is contraindi-

cated in situations where minimally invasive surgery is contraindicated.

WarningS and prEcauTiOnS:

 1) The Endoscope is a reusable instrument that is supplied 

non-sterile. Thoroughly clean and sterilize the Endoscope prior to each use. Follow recommended 
cleaning and sterilization instructions. 2) The Endoscope is a fragile instrument. Handle carefully 
to avoid breakage. Bending of the Shaft or dropping the Endoscope may damage the optics or 
other internal components, making the Endoscope inoperable. Store in a protective tray whenever 
possible. 3) The Dissection Tip is sterile unless the package is opened or damaged. The Tip is de-
signed for single use. Do not reuse or resterilize the Dissection Tip. 4) Whenever using endoscopic 
illumination equipment, ensure that the light cable connectors never rest on flammable materials 
such as surgical drapes, towels, etc. 5) If Prevacuum or Gravity steam sterilization (autoclave) are 
used, the Endoscope may have a shorter life due to harsher sterilization environment. Inspect the 
Endoscope after each steam sterilization cycle for damage. 6) Do not cool a hot Endoscope after 
sterilization by rapidly exposing it to air or liquid. Sudden temperature changes may cause glass 
components to crack. 7) Do not “flash” steam sterilize (autoclave) the Endoscope. Flash (i.e., 
unwrapped) steam sterilization cycles introduce sudden temperature changes, which may cause 
glass components to crack.

This document is not intended to replace the Instructions for Use. Please refer to the full Instruc-
tions for Use for more detailed information.

Содержание VASOVIEW 7

Страница 1: ...VASOVIEW 7 Training Manual English v1 Copyright MAQUET Gmbh Training Manual VASOVIEW 7 Endoscopic Vessel Harvesting System...

Страница 2: ...t judgement of the medical professional performing the procedure VASOVIEW 7 Training Manual MAQUET has a strong history pioneering and advancing endoscopic vessel harvesting systems The VASOVIEW 7 xB...

Страница 3: ...on 13 Branch Cauterization and Transection 16 Optimizing Visualization 19 Guidelines for Safe and Effective Use 20 BiSECTOR Bipolar Ligating Forceps Running the Vein 21 Distal Transection and Ligation...

Страница 4: ...nsecting Branches Insert the VASOVIEW 7 Harvesting Cannula to cauterize and transect branches using the BiSECTOR Bipolar Ligating Forceps Enhanced Visualization For enhanced visualization during vesse...

Страница 5: ...the radial artery pedicle and expose the branches Using the BiSECTOR tool cauterize and transect branches Running the Vessel Deploy the C Ring from the VASOVIEW 7 Harvesting Cannula engage the radial...

Страница 6: ...Video Monitor Light Source Insufflator Insufflator tubing CO2 Source Fiber Optic Cable with appropriate Scope Adaptor End Camera compatible to Endoscope 7 mm Extended Length Endoscope 7 mm Endoscope...

Страница 7: ...t should be checked for damaged fiber optics To do this hold one end toward a direct light source and look into the other end If it illuminates with no more than one third of the area grayed the cable...

Страница 8: ...d the knee area Once the intended incision site is identified mark the area with a surgical marker The saphenous vein can usually be found by using anatomical landmarks For morbidly obese patients an...

Страница 9: ...tor etched on the endoscope shaft Figures 2 and 3 Insert the endoscope with attached Dissection Tip through the BTT Ensure that the Endoscope Seal is on the BTT it comes packaged with this seal attach...

Страница 10: ...dyle Make a 2 cm incision along the posterior border of the tibial epicondyle Figure 7 Above the knee Make a 2 cm incision in the thigh where the vein lies in the groove between the sartorius and grac...

Страница 11: ...nsert the Dissection Tip into the space created and place it onto the anterior surface of the saphenous vein Note on the video monitor the color of the vein and the surrounding yellow fatty tissue It...

Страница 12: ...d can be used to inflate the balloon on the BTT Inflate the balloon with the minimal amount of air needed to create an adequate seal 0 25 cc s of air If there is obvious leakage around the incision us...

Страница 13: ...flation aids the dissection by enabling constant visualization reduces bleeding and helps maintain a working space Mastery of the dissection will enable clear exposure and easy access for branch caute...

Страница 14: ...anipulation If tissue collects on the Dissection Tip use your free hand on the outside of the leg to palpate the tip and dislodge the tissue Figure 16 Maintain the endoscope in a position parallel to...

Страница 15: ...sue that is still adhered Figure 18 When a branch is encountered dissect it with short controlled gentle probes with the Dissection Tip Con tinue until all branches are adequately exposed to facilitat...

Страница 16: ...handle of the VASOVIEW 7 Harvesting Cannula The endoscope is secured with an audible click locking it into place Figure 21 Attach the Bipolar Cord to the VASOVIEW Harvesting Cannula Ensure that the Bi...

Страница 17: ...de Rotate the BiSECTOR tool manually to orient the blade upward for ease of visualization and branch capture Figures 23a 23b When the branch is captured rotate the BiSECTOR tool away from the vein the...

Страница 18: ...my to further expose the branch Set the bipolar generator wattage according to the type of bipolar generator being used Refer to the VASOVIEW 7 xB Endoscopic Vessel Harvesting System Instructions for...

Страница 19: ...of the Harvesting Cannula If the distal lens of the endoscope becomes obscured by blood or fat use the integrated endoscope wash feature Fill the 5 cc syringe with saline and attach it to the blue Sc...

Страница 20: ...vessel branch cauteri zation and transection working back towards the incision If branches prohibit advancement of the Harvesting Cannula it may be less traumatic to the vein to perform vessel branch...

Страница 21: ...f the vein Externalize the vein through the stab wound and divide it under direct vision per surgical protocol Figure 30 Allow the harvested vein segment to retract back into the tunnel Ligate the ter...

Страница 22: ...f the vein for vessel preparation Measure the vein to ensure adequate length for the bypass grafting Figure 31 If additional length is needed further dissection can be carried out through the same inc...

Страница 23: ...assembly information see the Product Assembly section of the Endoscopic Vessel Harvesting of the saphenous vein portion of this manual page 9 Attach the 7 mm Extended Length Endoscope to the camera co...

Страница 24: ...under wrist Note the time of tourniquet deployment Generally tourniquet time is kept under one hour ENDOSCOPIC DISSECTION The radial artery is dissected as a pedicle with the accompanying venae comit...

Страница 25: ...Endoscope along the posterior aspect of the radial artery again avoiding contact with the radial artery Although anatomy varies from patient to patient posterior branches are likely to be encountered...

Страница 26: ...e settings differ by type of generator The fascia should be released from the distal forearm to the mid or proximal forearm Figure 39 BRANCH CAUTERIZATION AND TRANSECTION Once the fasciotomy is comple...

Страница 27: ...ures without a tourniquet the radial pedicle should be clamped at the wrist with a soft bulldog type clamp prior to proximal division This is intended to keep the pedicle from bleeding into the tunnel...

Страница 28: ...onal hemostasis Procedure Using a Tourniquet Deflate the tourniquet and note the tourniquet time Check pulse oximetry on the digits to ensure adequate perfusion to the hand The radial artery is prepar...

Страница 29: ...SOVIEW 7 Endoscopic Vessel Harvesting System kit to identify the location and the length of the vessel harvested A sticker is in cluded for the patient s hospital record and a card is included for the...

Страница 30: ...primary troubleshooter once the procedure has begun See Troubleshooting Section Turn on all components to the video system including those that may not be used Once the various cords are handed off fr...

Страница 31: ...vein harvesting is completed and to simplify the general set up Sterile items Components of the VASOVIEW 7 xB EVH System kit VASOVIEW Harvesting Cannula with BiSECTOR tool and C Ring Short Port BTT w...

Страница 32: ...fragile than shorter or larger diameter endoscopes It is recommended that the endoscopes be kept in the instrument tray available through MAQUET Cardiovascular More detailed instructions can be found...

Страница 33: ...on tunnel wall Remove scope and clean it externally Use scope washer within the Harvesting Cannula to clean the lens Endoscope Fogging Use anti fog solution on the end of the endoscope only with the...

Страница 34: ...both suture ends be tween the thumb and index finger as shown Step 2 Take the shortest suture tail in front of the 2 sutures forming a complete circle around the index finger Step 3 Take the shortest...

Страница 35: ...OVIEW 7 Training Manual English v1 Copyright MAQUET Gmbh Step 5 Pull short and long strand to tighten knot Step 6 Pull short strand and one of the 2 strands to tighten knot securely Remember one of th...

Страница 36: ...the knot is secure cut the short tail leaving 2 to 3 mm of suture Do not cut right next to the knot Step 8 Thread the knot pusher and advance knot and knot pusher into insufflated tunnel Push the knot...

Страница 37: ...osition 16 When performing radial artery harvesting the radial artery harvesting procedure should be performed prior to placing the patient on cardiopul monary bypass 17 Do not touch the coagulating s...

Страница 38: ...38 ENDOSCOPIC VESSEL HARVESTING MAQUET Medical Systems USA VASOVIEW 7 Training Manual English v1 Copyright MAQUET Gmbh...

Страница 39: ......

Страница 40: ...s on patient mobility and wound management solutions GETINGE pro vides solutions for infection control within healthcare and contamina tion prevention within life sciences MAQUET specializes in soluti...

Отзывы: