
0
OUT
IN
OUT
IN
KMP-013aX
Connections
1.
Inlet
2.
Outlet
3.
Drainage bottle
4.
Condensation drain valve
Empty the air reservoir by pressing the
pin on the drain valve while the
compressor is running.
-
Compressed air
When the Standby function is required, air
from the central compressed air supply is
connected to the inlet.
The compressed air hose to the ventilator is
connected to the compressed air outlet
Drainage bottle
There is a water separator in the compressor.
It is automatically emptied twice a minute by
a solenoid valve. The water collected in the
water separator drains into the drainage
bottle. A brief ”pssh” sound is heard each time
the solenoid valve opens. The bottle must be
checked regularly and emptied when
necessary. The higher the degree of humidity
in the ambient air, the quicker the bottle fills
Condensation drain valve
The drain valve is used to ensure that there is
no water build-up in the air reservoir by
releasing any collected condensed moisture.
If the compressor is running continuously, a
small amount of water may collect in the air
reservoir.
10
COMPRESSOR MINI, User´s Manual
English version
28.0
2. Device description
Содержание COMPRESSOR MINI
Страница 1: ...User s Manual VENTILATION ANESTHESIA COMPRESSOR MINI...
Страница 2: ......
Страница 4: ......
Страница 8: ...6 COMPRESSOR MINI User s Manual English version 28 0 1 Before use...
Страница 14: ...12 COMPRESSOR MINI User s Manual English version 28 0 3 Preparation Connection...
Страница 18: ...16 COMPRESSOR MINI User s Manual English version 28 0 5 Preventive maintenance Storage...
Страница 20: ...18 COMPRESSOR MINI User s Manual English version 28 0 6 Cleaning...
Страница 22: ...20 COMPRESSOR MINI User s Manual English version 28 0 7 Alarms...
Страница 24: ...22 COMPRESSOR MINI User s Manual English version 28 0 8 Troubleshooting...
Страница 27: ......