background image

Mantracourt Electronics Limited

 – ICA User Manual   

 

11 

Chapter 3 Calculating Offset and Gain Resistor Values 

Calculating the offset resistor values 

The ICA1H and 2H can be used in a pseudo ‘bipolar ±’ mode by fitting a resistor ‘R

offs

’ as shown in Figure 2.2. This will 

shift the output voltage to a known level (V

offs

) when the input is zero (0 mV/V) allowing both tension and compression 

of the load cell to be measured. 
Use the following formula for the ICA1H and 2H to calculate the value of R

offs

 in kΩ. 

 

𝑅

𝑜𝑓𝑓𝑠

=

148

(𝑉

𝑜𝑓𝑓𝑠

− 0.1)

 

 
The gain will also need to be changed to account for smaller changes in output resulting from a given input change. 
This means that the total mV/V will change: 
For the ICA1H: 

𝑟𝑒𝑞𝑢𝑖𝑟𝑒𝑑 mV/V =

10V

𝑏𝑖𝑝𝑜𝑙𝑎𝑟 𝑜𝑢𝑡𝑝𝑢𝑡 𝑐ℎ𝑎𝑛𝑔𝑒

  × 𝑙𝑜𝑎𝑑 𝑐𝑒𝑙𝑙 mV/V

 

 

For the ICA2H: 

𝑟𝑒𝑞𝑢𝑖𝑟𝑒𝑑 mV/V =

5V

𝑏𝑖𝑝𝑜𝑙𝑎𝑟 𝑜𝑢𝑡𝑝𝑢𝑡 𝑐ℎ𝑎𝑛𝑔𝑒

  × 𝑙𝑜𝑎𝑑 𝑐𝑒𝑙𝑙 mV/V

 

 

 

Calculating the gain resistor values 

The ICA range of strain gauge amplifiers are supplied un-calibrated but optimised for a sensitivity of 2.5 mV/V. To 
accommodate other sensitivities the gain resistor ‘R

gain

’ shown in the connection diagrams above can be changed 

according to the following formulae. 
N.B. a high quality component e.g. 1 % 10-15 ppm/°C metal film resistor should be used for optimum performance. It 
may be necessary to use a value from the less common E96 series to optimise the trim range. 
 

𝐺𝑎𝑖𝑛 = (

Gain constant

𝑟𝑒𝑞𝑢𝑖𝑟𝑒𝑑 𝑚𝑉/𝑉

)                                 𝑅

𝑔𝑎𝑖𝑛

= (

Resistor constant

(𝐺𝑎𝑖𝑛 − 1)

) Ω

 

 

  

Where: 
 
Model 

Output range 

Gain constant 

Resistor constant 

ICA1H 

0.1 to 10.1 V 

372 

40000 

ICA2H 

0.1 to 5.1 V 

186 

40000 

ICA3H 

± 10 V 

372 

40000 

ICA4H 

4 to 20 mA 

160 

63200 

ICA6H 

± 10 V 

372 

40000 

 
 
The ICA4H has two surface mount gain resistors. These do not need to be removed if the gain is to be increased (a 
lower mV/V input level). However, they need to be taken into account and an extra calculation stage is required: 

 

 

𝑅 = (

1

(1/𝑅

𝑔𝑎𝑖𝑛

) − 0.001

) Ω

 

Содержание ICA1H

Страница 1: ...ICA User Manual ICA ICA Embedded Strain Gauge Analogue Amplifiers...

Страница 2: ...nt Resistance Formula ICA4 and ICA5 9 Table 2 1 The ICA H S Cable Data 10 Chapter 3 Calculating Offset and Gain Resistor Values 11 Calculating the offset resistor values 11 Calculating the gain resist...

Страница 3: ...5 mV V and greater than 10 mV V are possible As supplied they are optimised to 2 5 mV V This range covers most but not all strain gauges Sensitivity adjustment SPAN is achieved by a combination of ch...

Страница 4: ...ure and protected Figure 2 1 Dimensions The module is designed to fit in the strain gauge pocket Use the 2 1mm hole to secure the unit The mounting hole will accept an M2 screw or American equivalent...

Страница 5: ...r the ICA1H ICA2H The strain gauge cable should be attached to the solder pads as illustrated For most applications 3 wire un shielded field wiring is quite adequate For best EMC performance use the c...

Страница 6: ...e cable should be attached to the solder pads as illustrated For most applications 3 wire un shielded connections for field wiring is quite adequate For best EMC performance use the connections shown...

Страница 7: ...onitor connected to the loop see the Output Shunt Resistance Formula at the end of this chapter Figure 2 7 Connection Details for the ICA4H Source mode Figure 2 8 Connection Details for the ICA4H Sink...

Страница 8: ...7 The ICA4H can be used with three wire cabling in both sink and source mode The only difference between the two modes is whether the common end of the load is connected to the positive of the PSU si...

Страница 9: ...ower signal level requires more gain from the ICA5 to compensate leading to a degradation in performance with regards to temperature stability and noise performance Impedances greater than 1000 ohms a...

Страница 10: ...luding the wiring resistance for low supply voltage operation of the ICA4H and ICA5 Alternatively by transposing the formula the minimum supply voltage for a given shunt resistance can be determined I...

Страница 11: ...K Farnell 585 646 Individually shielded twisted multi pair cable 7 0 25 mm 3 pair Tinned copper drain Individually shielded in polyester tape Diameter 8 1 mm Capacitance m core to core 98 pF core to s...

Страница 12: ...in resistor values The ICA range of strain gauge amplifiers are supplied un calibrated but optimised for a sensitivity of 2 5 mV V To accommodate other sensitivities the gain resistor Rgain shown in t...

Страница 13: ...1 40000 37 2 1 1105 Use preferred value 1k1 Example 3 ICA3H no offset 2 5 mV V required sensitivity at 5 V output Gain constant Required output Maximum output 372 2 5 5 10 74 5 Resistor constant 1 40...

Страница 14: ...s the sensitivity of the load cell in mV V Vexc is the excitation voltage in volts and Z is the load cell impedance in ohms e g For a 2 5 mV V 1000 ohm load cell R2 222 17 k ohms use 220 k nearest E24...

Страница 15: ...l 14 Offset Resistor R1 vs Load Cell Impedance The following table gives the value of R1 for various load cell impedances and 2 FS and 4 FS trim Load Cell Impedance 2 FS 4 FS 350 Ohms 30k n a 700 Ohms...

Страница 16: ...Details for Calibration 1 Apply the known low calibration conditions weight force or mV V This may be zero if required and using the Z potentiometer Zero set the output to the relevant low level depe...

Страница 17: ...ould be high and fixed a Check the sensor is connected typically reading 350 ohm across output with the power off b Check the Excitation voltage is at 5 V dc for the ICA1 4H and 6H see Table 3 6 for I...

Страница 18: ...a If the adjustment cannot reach the maximum output desired then check the tare is not too high b If the potentiometer does not alter the output the unit must be repaired remove from service c It is...

Страница 19: ...overdue accounts and that a strict interpretation of our conditions of trading invalidates warranty claims where late payment has occurred Please refer to the RMA Form Return Material Authorization co...

Страница 20: ...dentified as SPAN It relates to the proportional output to the sensor input Calibration of the ICA is determined by setting the Gain Span and Offset Zero The amount of amplification used in an electri...

Страница 21: ...force exerted on it be it pressure torque or load The ICA is designed to convert this change in the resistance of the Strain Gauge to a proportional electrical signal Strain Gauge Amplifier The ICA is...

Страница 22: ...tal current 22mA Measurement Parameter Minimum Typical Maximum Units Notes Bridge Excitation 4 90 5 5 10 volts Bridge Impedance 350 1000 5000 ohms Bridge Sensitivity 0 5 2 5 50 mV V Note 1 Output load...

Страница 23: ...surement Parameter Minimum Typical Maximum Units Notes Bridge Excitation 4 90 5 5 10 volts Bridge Impedance 350 1000 5000 ohms Bridge Sensitivity 0 5 2 5 50 mV V Note 1 Output load 5000 ohms Bandwidth...

Страница 24: ...ent Parameter Minimum Typical Maximum Units Notes Bridge Excitation 4 90 5 5 10 volts Bridge Impedance 350 1000 5000 ohms Bridge Sensitivity 0 5 2 5 50 mV V Note 1 Output load 5000 ohms Bandwidth DC 1...

Страница 25: ...rement Parameter Minimum Typical Maximum Units Notes Bridge Excitation 4 90 5 5 10 volts Bridge Impedance 350 1000 5000 ohms Bridge Sensitivity 0 5 2 5 150 mV V Note 1 Output load 900 ohms Note 2 Band...

Страница 26: ...stment 8 FR Linearity 0 02 FR Temperature stability Zero Temperature Stability 0 001 0 005 FR C At 2 5 mV V Span Temperature Stability 0 007 0 014 FR C At 2 5 mV V FR Full Range 16 mA Note 1 1000 ohm...

Страница 27: ...1000 Hz Zero adjustment 2 FR Note 4 Span adjustment 8 FR Linearity 0 02 FR Temperature stability Zero Temperature Stability 0 001 0 005 FR C At 2 5 mV V Span Temperature Stability 0 007 0 014 FR C At...

Страница 28: ...ny warm up time Note 2 Not including excitation current e g when connected to a 350 ohm load cell excitation current 5 350 14 mA Total current 22 mA Measurement FR Full Range 10 V Note 1 Set by calibr...

Страница 29: ...nformity with the following relevant Union harmonisation legislation LVD directive 2014 35 EU EMC directive 2014 30 EU RoHS directive 2011 65 EU Based on the following harmonised standards EN 61326 1...

Страница 30: ...Mantracourt Electronics Limited ICA User Manual 29...

Страница 31: ...Mantracourt Electronics Limited ICA User Manual 30...

Страница 32: ...Manual Applies To ICA Product Range Part Number 517 936 Issue Number 01 00 Dated 19th January 2018 In the interests of continued product development Mantracourt Electronics Limited reserves the right...

Отзывы: