24
CG-2 | 1.04.00
www.mc-techgroup.com
17.1
CHECK OF THE EFFICIENCY
To check the efficiency of the converter cartridge a test gas with a NO
2
-concentration according to the
concentration in the sample gas (e.g. 20 ppm in N
2
) is necessary.
Feed test gas upstream the converter and read measured NO-concentration at the analyzer.
Measured NO-concentration [ppm]
Efficiency = ------------------------------------------------- x 100 %
NO
2
-concentration in test gas [ppm]
For a precise result the measuring range of the analyser has to be adapted to the concentration of the
test gas!
17.2
CONVERSION PPM --> MG/M
3
1 ppm --> mg/m
3
DIN 1343
NTP
STP
SATP
T=0 °C
T=20 °C
T=25 °C T=25 °C
p=1013 mbar p=1013
p=1000
p=1013
Nitrogen monoxide NO
1.34 mg/m
3
1.25
1.21
1.23
Nitrogen dioxide NO
2
2.05 mg/m
3
1.91
1.86
1.88
Standard conditions
DIN 1343
Standard conditions in Germany
NTP
Normal temperature and Pressure mainly in the USA
STP
Normal conditions weltweit standardized
SATP
new Normal conditions
“Standard Ambient Temperature and Pressure“
The table is based on the following calculation. This calculation is based on the ideal gas laws
Molecular weight NO = 30.01
Molecular weight NO
2
= 46.0055
𝑚𝑔
𝑚³
=
0.1 𝑥 𝑀𝑜𝑙𝑒𝑐𝑢𝑙𝑎𝑟 𝑤𝑒𝑖𝑔ℎ𝑡 𝑥 𝑃𝑟𝑒𝑠𝑠𝑢𝑟𝑒 𝑥 𝐶𝑜𝑛𝑐𝑒𝑛𝑡𝑟𝑎𝑡𝑖𝑜𝑛 𝑖𝑛 𝑝𝑝𝑚
𝑀𝑜𝑙𝑒𝑐𝑢𝑙𝑎𝑟 𝑔𝑎𝑠 𝑐𝑜𝑛𝑠𝑡𝑎𝑛𝑡 8.314472
𝐽
𝑘𝑚𝑜𝑙𝐾
𝑥 𝑅𝑒𝑓𝑒𝑟𝑒𝑛𝑐𝑒 𝑡𝑒𝑚𝑝𝑒𝑟𝑎𝑡𝑢𝑟𝑒 𝑖𝑛 𝐾𝑒𝑙𝑣𝑖𝑛
Содержание CG Series
Страница 1: ...Gas Converter CG CG 2 2M CG 2H W 2MH W Instruction Manual Version 1 04 00 ...
Страница 8: ...8 CG 2 1 04 00 www mc techgroup com Hot surface Contact may cause burn Do not touch ...
Страница 33: ...www mc techgroup com CG 2 1 04 00 33 Figure 11 Wiring plan CG 2M Drawing No 2224 5 04 0 ...
Страница 34: ...34 CG 2 1 04 00 www mc techgroup com Figure 12 Wiring plan CG 2MH W Drawing No 2224 5 04 4 ...
Страница 35: ...www mc techgroup com CG 2 1 04 00 35 Figure 13 Wiring plan CG 2H W Drawing No 2224 5 04 5 ...