Step 1
Please read the full “Instructions For Use” before using
this product. Grasp the probe body and align the tab
with the holster notch. Push the probe base into the
holster while maintaining the alignment.
Push the active biopsy button only once to initiate the
biopsy cycle. To interrupt the biopsy cycle, push the biopsy
button again when the cutter is retracting from position
1 to 3 (see image below) or when the cutter is advancing
from position 3 to 2. The cutter will retract to position 3,
and the biopsy indicator light will flash orange. Complete
the cycle by pushing the biopsy button.
Push the active biopsy button on the holster to acquire
a sample. The biopsy cycle proceeds: a vacuum is
created, the inner cutter of the probe is retracted,
tissue is drawn into the aperture and the cutter rotates
and translates forward, cutting tissue.
When the cutter reaches the distal end of the aperture,
the vacuum will transport the sample to the specimen
collection cup.
Step 2
Rotate to desired position and release.
Step 1
Push the needle rotation knob in the direction shown
below to rotate.
Disconnect the integrated coaxial cannula and
remove the probe by pressing the coaxial release
button, leaving the cannula in place.
Interrupt a biopsy cycle (as previously described)
leaving the aperture open. Inject 1cc of sterile saline
into the needle aperture. Complete the biopsy cycle by
pushing the biopsy button. The system is now primed
and ready to sample.
Needle Rotation
Biopsy Cycle Interrupt
Biopsy Tissue
Removing the Coaxial Cannula
Priming the System
Step 1
Ensure the probe is attached. Press and hold the two
inactive biopsy buttons simultaneously for three seconds.
There will be an audible indicator and all three indicator
lights will flash.
Step 2
Press and release the desired active biopsy button. The
active biopsy button is stored in the device for future use.
Step 2
Push the probe firmly into the holster until a click is
heard, indicating the locking tabs have engaged the
holster.
Note: Automatic initialization will begin.
Assembly
Biopsy Button Selection