Step 1

Please read the full “Instructions For Use” before using 

this product. Grasp the probe body and align the tab 

with the holster notch. Push the probe base into the 

holster while maintaining the alignment.

Push the active biopsy button only once to initiate the 

biopsy cycle. To interrupt the biopsy cycle, push the biopsy 

button again when the cutter is retracting from position 

1 to 3 (see image below) or when the cutter is advancing 

from position 3 to 2. The cutter will retract to position 3, 

and the biopsy indicator light will flash orange. Complete 

the cycle by pushing the biopsy button.

Push the active biopsy button on the holster to acquire 

a sample. The biopsy cycle proceeds: a vacuum is 

created, the inner cutter of the probe is retracted, 

tissue is drawn into the aperture and the cutter rotates 

and translates forward, cutting tissue. 

When the cutter reaches the distal end of the aperture, 

the vacuum will transport the sample to the specimen 

collection cup.

Step 2

Rotate to desired position and release.

Step 1

Push the needle rotation knob in the direction shown 

below to rotate.

Disconnect the integrated coaxial cannula and 

remove the probe by pressing the coaxial release 

button, leaving the cannula in place.

Interrupt a biopsy cycle (as previously described) 

leaving the aperture open. Inject 1cc of sterile saline 

into the needle aperture. Complete the biopsy cycle by 

pushing the biopsy button. The system is now primed 

and ready to sample.

Needle Rotation

Biopsy Cycle Interrupt

Biopsy Tissue

Removing the Coaxial Cannula

Priming the System

Step 1

Ensure the probe is attached. Press and hold the two 

inactive biopsy buttons simultaneously for three seconds. 

There will be an audible indicator and all three indicator 

lights will flash.

Step 2

Press and release the desired active biopsy button. The 

active biopsy button is stored in the device for future use. 

Step 2

Push the probe firmly into the holster until a click is 

heard, indicating the locking tabs have engaged the 

holster.  
Note: Automatic initialization will begin.

Assembly

Biopsy Button Selection