- 21 -
•
Prior to using the device, check the milk bottle for cracks and exchange it if need be. A frequent boiling
of the milk bottle and the parts of the suction kit may cause a brittling of the plastic material resulting in
cracks. Porous parts shall be replaced by new parts.
•
Small parts are enclosed to the product which could be inhaled or swallowed. Keep them away from
toddlers/children.
5.1 Note on electromagnetic compatibility (EMC)
The
mamivac
®
SENSITIVE-C
breast pump described in these instructions for use may only be operated
in accordance with the following information, provided that the accessories and spare parts mentioned
in chapters 11 and 14 are used, in an electromagnetic environment that corresponds to a domestic environ-
ment (dwelling, medical practices) where the public power system is available.
It is not permitted to operate the
mamivac
®
SENSITIVE-C
breast pump in means of transportation, be
it over land, across water or through the air, nor in special-use HF shielded rooms (e.g. MRT) where elec-
tromagnetic bursts of high intensity may occur nor in the immediate vicinity of active HF surgery devices or
shortwave therapy devices.
A detailed technical description of the electromagnetic environment can be found in the enclosed
EMC guidelines.
Provided that the
mamivac
®
SENSITIVE-C
breast pump is operated in the specified environment, an im-
pairment of the performance characteristics given in chapter 11 or the basic safety is not to be expected.
WARNING:
Avoid using this device directly next to other devices or with other devices stapled on top of
each other as this could disturb operation.
WARNING:
The use of accessories, converters or cables other than those defined or provided by Kimetec
for this device can lead to an increased electromagnetic interference or a reduced electromagnetic immu-
nity of the device resulting in faulty operation.
WARNING:
Portable HF communication devices (transmitters), including their accessories such as
antenna cables and external antennas, shall not be used within a distance shorter than 12 inches (30 cm)
to the parts and cables listed by Kimetec for the
mamivac
®
SENSITIVE-C
breast pump. Non-observance
may impair the performance characteristics of the device.
6. Storage, transportation and ambient conditions
•
Avoid strokes or impacts of any kind, regardless of whether the device is packed or not.
•
Avoid rapid temperature changes, high air humidity or a dusty environment.
•
The device can be operated at an ambient temperature of +5 °C to +40 °C, a relative humidity of
15 % to 93 % (non-condensing) and an air pressure of 700 hPa to 1060 hPa.
•
Store and transport the device at an ambient temperature of -29 °C to +60 °C, a relative humidity
of max. 85 % and an air pressure of 500 hPa to 1060 hPa.
•
An excessive accumulation of lint and dust in the inside of the device impairs heat dissipation of
components, which may overheat as a result representing a fire hazard.
7. Warranty
The warranty period for the
mamivac
®
SENSITIVE-C
breast pump is 36 months from the date of purchase.
Improper use, opening the device or using suction kits not specified for this breast pump will void
any warranty claims.