The following steps should be taken when storing clinical beds.
Pull out the mains plug and keep safe for trans-
port.
Supply the battery (optional) regularly with power
so that deep discharge does not occur.
Remove any accessories such as bed lamps,
trapeze bars, etc.
Cover the clinical bed so that the head/foot
boards and the frame cannot be damaged.
Mark the storage date clearly on the bed (due to
maintenance intervals)
Lock the hand controller
Caution! The same conditions apply to the stor-
age of clinical beds as to the working environ-
ment (temperature, humidity, heat, etc.)
The manufacturer’s transport aid should be
used to transport the beds.
The manufacturer is only liable for the safety and reliability of
the product if it is maintained regularly and used in accord-
ance with the safety instructions. If any significant faults are
found during maintenance work which mean the safe opera-
tion of the product cannot be guaranteed, the product should
be taken out of use. Maintenance work should be carried out
at least once a year.
The regulations within the general Operator Ordinance and
the regulations for use of electrically movable devices §6 MP-
BetreibV as well as the BGV A3 – electrical units – and equip-
ment in accordance with BetrSichV, TRBS, DIN VDE 0701-
0702 apply.
We offer inspection plans for servicing our clin-
ical beds in a separate service guide!
Any faults which impair the function and
safety of the clinical bed must be re-
solved before the bed is used again and
must be reported to the relevant person.
Only original parts from Betten Malsch GmbH
should be used.
M A I N T E N A N C E