15
Operation stops with
alarm:
Display shows
Over Voltage
- Verify that the running buffer is correct.
- Verify the all cables are attached correctly
- Turn the Power switch off and on again;
restart application.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord from
the outlet, and contact Technical Service.
Operation stops with
alarm:
Display shows
Ground leak detected
during run
Check the electrophoresis system for
improper grounding. Restart the power supply
by turning the Power switch off and on.
Operation stops with
alarm:
Display shows
Power supply is
overheating
- Turn off power supply. Check for sufficient
airflow around the power supply fan. After
cooling down, restart the power supply by
turning the Power switch to the on position.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord from
the outlet, and contact Technical Service.
Operation stops with
alarm:
Display shows
Bad connections for
terminal connectors or
damaged wires or
damaged platinum wires
Check all the connections to terminators,
cables, wires, and gel tanks
Operation stops with
alarm:
Display shows
and
countdown
Power supply is getting
restart ON
- Verify that the electrophoresis condition and
system is correct.
- Pay attention to the own safety.
Encountering Problems
1. Check the troubleshooting section.
2. Call Technical Service or e-mail to
3. If the unit must be shipped back for repair, contact Major Science or the
distributor for a Return Authorization Number and shipping instructions. The
unit will be repaired as quickly as possible and returned to you.