4.2.4 Establishing a PC-Connection
To transfer data to a PC, establishing a PC-connection via USB is required. If the
touchTymp is used with office equipment that is not a medical device itself (see Table
4), make sure to establish the PC-connection in one of the following ways (see Table
4, PC Connection 2, 3 or 4).
Make sure you use only office equipment with the device
that is a medical device itself or meets the requirements of
IEC 60950. If a non-medical device is used within the patient
environment (1.5 m from patient as defined in IEC 60601) a
voltage transformer must be used (exception: a battery
driven laptop is used).
Table 4 PC-Connections
PC CONNECTIONS
PC Connection 1:
Medical device
– Medical Device
PC Connection 2:
Medical device
– Non-Medical Device
PC Connection 3:
Medical device
– Non-Medical Device
PC Connection 4:
Medical device
– Laptop (battery-driven)
4.2.5 Storage
When the touchTymp is not in use, store in a location where it will be safe from damage
to the touchscreen or other sensitive components such as the acoustic transducers
and cables. Store according to the recommended temperature conditions described in
section 6.1.
WARNING