ITEM
EXPLANATION
Frequencies
Press
Enter
to access this submenu. Select/deselect the
frequencies that are active and selectable during the test (Figure 23
and Figure 24).
Figure 23
Figure 24
NOTE
: 1000 Hz cannot be turned off and is therefore excluded from the list.
Signal
Select the default tone type of
Steady
or
Pulse
. Steady presents
the signal continuously during the tone presentation.
Pulse
turns
signal on and off every 250 ms during tone presentation.
Return to
1 kHz
After having reached the highest (i.e. 8 kHz) or lowest (125 Hz) test
frequency device returns to 1 kHz automatically. When set to
Off
,
it will continue to next highest/lowest frequency.
Level
increase
This function sets the direction of the
arrow
▲ ▼ keys to increase
the hearing level. Choose between:
•
Increase up
: Pressing ▲ will increase the hearing level (ie. 35 dB
to 40 dB). Recommended to use with
Display type
Table
.
•
Increase down:
Pressing the ▼ will increase the hearing level (ie.
35 to 40 dB). Recommended to use with
Display type
Graph
.
5.5.5
Start with…
Figure 25
Select the test selected upon turning on the
device:
PILOT TEST
or
Audiometry
5.5.6 Display Language
Figure 26
Select the operating language of the device
(
26).
Display
languages
include: