8 Follow up measurements
If your MB 11 system was purchased with the advanced features of the
Standard ABR or Time Step Stimulus, you can enable these functions by
clicking on the box labeled "advanced" on the lower right of the start screen.
The start screen now displays two additional tests: “Time Step Stimulus” and
“Standard ABR”.
These two advanced tests should only be used by the experienced
expert!
This offers the possibility of an immediate advanced evaluation of
hearing after a screening result of “REFER” has occurred. For this purpose,
you can choose between Standard ABR with adjustable intensity or the Time
Step Stimulus method which tests six intensities simultaneously.
Under good test conditions it is possible for an advanced tester to perform
these measurements with a calm or sleeping baby. If you are not able to get
good test results you should keep the baby calm by possibly feeding or
sedating it. You could also consider testing the baby another time when the
baby is sleeping.
IMPORTANT NOTE
The electrode placement of Vertex/Nape of the neck that is recommended
for the screening is not ideal for the Standard ABR or Time Step Stimulus
response when visualization of the Wave I of the ABR is desirable. Expert