Operating Instructions MA 42
Operating_Instructions_8100460-6_MA42_e_15a
61
8100460-6 1/15
Calibration values:
Sound field (0 degree incidence)
Reference equivalent threshold sound pressure level and maximum hearing
levels
For Canton CD 220 / Maico SBC speaker
ISO 389 – 7 and ANSI S3.6-1996
Signal
IEC60645-2
RETSPL [dBSPL]
IEC Max Level
[dBHL]
CD220 / SBC
ANSI S3.6
RETSPL [dBSPL]
ANSI Max Level
[dBHL]
CD220 / SBC
Speech
0.0
90 / 80
14.5
90 / 80
SN
0.0
90 / 75
14.5
90 / 75
WN
0.0
90 / 75
0.0
90 / 75
Frequency [Hz]
Tone RETSPL
[dBSPL]
NBN RETSPL
[dBSPL]
Max level Tone
[dBHL]
CD220 / SBC
Max level NBN
[dBHL]
CD220 / SBC
125
22.0
22.0
65 / -
55 / -
250
11.0
11.0
75 / -
70 / -
500
4.0
4.0
85 / 85
75 / 75
750
2.0
2.0
85 / 85
75 / 75
1000
2.0
2.0
85 / 85
75 / 75
1500
0.5
0.5
90 / 85
80 / 75
2000
-1.5
-1.5
90 / 85
85 / 75
3000
-6.0
-6.0
90 / 90
85 / 80
4000
-6.5
-6.5
95 / 90
85 / 80
6000
2.5
2.5
85 / 90
80 / 80
8000
11.5
11.5
80 / 75
70 / 65