Service Manual MA 42
D-0125965 Rev. D
20
18/05/2022
For the software calibration a PC connection is required.
4.2.2 Prepare the MA42 Device for Calibration
1. Connect all requested parts including USB and Power Supply.
2. Press the button on the back to power on the device.
3. Wait for the device to completely boot up.
4. From the main screen Press and hold the
button to view the
button.
You will then see two options popping up:
5. When in the Settings Menu press and hold the store S button for about 4 seconds
to open the Service User Login Screen
Press and hold
button for
about 4 sec.
Use the Dail to change to the
button and select by pressing
the
button