5.5 MA 27e Special Functions
Talk Forward
: On the MA 27e, Talk Forward is activated by
holding down the
“Hearing Level dB” rotary wheel. Rotating the
dial while in the talk-forward mode will adjust the level of the talk-
forward to the patient.
F-Key Functions
: The following F-key functionalities can be
accessed by pressing the
“Frequency Hz” rotary wheel:
F1:
Delete all thresholds stored in the internal memory of the
MA 27e.
F2:
Store a Not Heard threshold point.
F3:
Display the L/R thresholds stored in the internal memory
of the MA 27e.