19
5.2 Changing Analog Magnification (M1) [S]
1) Simultaneously press and hold
M/OFF
(1), then quickly press and release
inch/
mm
(2) once to enable the setup mode. (
M1
icon appears in the lower left cor-
ner of the LCD) (Figure 3).
Figure 3 -
Entering the Set-Up Mode
2) Press and release
inch/mm
to select the desired Analog Magnification signified
by the X1, X5, X10, X50, or no icon (blank) - no scale displayed (Figure 4).
3) Press and release
M/OFF
until it returns to the gaging mode. No “
Mx
” icon is
displayed. Where “
x
” could be 1, 2, 3, 12, 23, etc.
Display Icon
± Full Scale
Grad Value
(in/mm)
(in/mm)
no icon
X1
0.040 / 1.00
0.0020 / 0.050
X5
0.020 / 0.20
0.0010 / 0.010
X10
0.010 / 0.10
0.0005 / 0.005
X50
0.002 / 0.02
0.0001 / 0.001
uMaxum
II
bar graph disabled