15
4.0 Mounting Gage for Setup
The
µMaxum II
indicator, normally stem mounted, can be mounted in a gage or fixture
using an optionally available back or adapter.
1) If the unit is
OFF
press and release the
ZERO/ON
button to turn the unit
ON
. If
unit is already
ON
, press and hold
ZERO/ON
button to enter TSP Mode.
2) Mount the indicator in the gage or stand.
3) Place the zero or nominal master in the gage or stand.
4) Mechanically adjust the indicator until the reading is at or near the desired
value/or spindle position. For the
µMaxum II
Comparator a zero setting value is
ideal.
5) Securely lock the Indicator into position.
6) Press and release the
ZERO/ON
button to set the indicator to the zero or preset
value.
You are now ready to take readings. The bargraph should either be blank (if Setup to be
blank) or a full bargraph with multiple graduations lit (if setup to be x1, x5, x10 or x50).
NOTE: A single grad segment traveling left and right indicates the Indicator is in TSP
Mode and not in gaging mode
.