TORQUE 10 

GRAPH DISPLAY 

 

www.magtrol.com 

35

 

9.  GRAPH DISPLAY 

 

Figure 43 Graph display 

 

Control 

Function 

Graph Legend 

Located to the upper left of the graph the legend expands as more 
plot  parameters  are  selected.  Each  plot  is  identified  with  an 
abbreviated  name  and  is  color  coded.  There  are  many  plot 
options  available  by  right-licking  on  the  plot  indicator  in  the 
legend. You can change color, line style, line width, etc. from the 
pop-up menu. 

Cursors 

You  may  create  single  or  multiple  cursors  on  the  graph.  Each 
cursor will have a numeric X-Y value shown in the cursor display 
table. The cursor can then be dragged along the plot with your 
mouse, or moved with the left and right cursor buttons. 

CREATE A CURSOR 
Right-click  in  the  cursor  window.  Select  Create  Cursor  and 
Single-Plot. 
Cursor  0  will  now  appear  in  the  plot  area  and  in  the  cursor 
window. The cursor attributes are the default settings. 
You  may  change  the  parameter  the  cursor  is  locked  to  by 
selecting Snap To and dragging to the new parameter. The color, 
style, point style, etc. of the cursor itself can be changed from the 
Attributes menu. 
Repeat this process to add more cursors to the graph. 

Y1,2,3,4,5 

Axis 

(None - n) 

Selects  the  parameter  to  plot  on  the  first  Y-axis.  The  available 
parameters are the ones selected for acquisition during the test. 

X Axis (None - n) 

Selects  the  parameter  to  plot  on  the  X-axis.  The  available 
parameters are the ones selected for acquisition during the test. 

Содержание TORQUE 10

Страница 1: ...TORQUE 10 TORQUE TRANSDUCER SOFTWARE USER MANUAL www magtrol com...

Страница 2: ...on of Magtrol is strictly prohibited TRADEMARKS National Instruments LabVIEW and NI DAQmx are trademarks of National Instruments Corporation PURCHASE RECORD Please record all model numbers and serial...

Страница 3: ...appropriate safety guards 3 Periodically check all connections and attachments 4 Always wear protective glasses when working close to rotating elements 5 Never wear necktie or baggy clothes when stand...

Страница 4: ...TORQUE 10 d www magtrol com...

Страница 5: ...2 MODEL 3411 DRIVER 5 3 2 1 LAN Configuration 6 3 3 MODEL 3410 DRIVER 8 4 TORQUE 10 EXECUTATION 9 4 1 MENU 9 4 1 1 File menu 9 4 1 2 Language menu 10 4 1 3 Help menu 10 5 SETUP CONFIGURATION 11 5 1 D...

Страница 6: ...gtrol com 7 3 RUN TEST IN HS DATA ACQUISITION MODE 27 7 4 EXAMPLE OF DATA FILE 29 8 DATA DISPLAY 31 8 1 EXPORT TO EXCEL FUNCTION 32 8 2 PRINT 33 9 GRAPH DISPLAY 35 10 CREATING TASK 37 10 1 TEMPERATURE...

Страница 7: ...he manual is as follows Chapter 1 INTRODUCTION Contains the general information Chapter 2 SOFTWARE INSTALLATION Provides information needed for the installation of the TORQUE 10 software Chapter 3 DRI...

Страница 8: ...ICATES A HAZARDOUS SITUATION THAT IF NOT AVOIDED COULD RESULT IN DEATH OR SERIOUS INJURY THIS INTRODUCES DIRECTIVES PROCEDURES PRECAUTIONARY MEASURES ETC WHICH MUST BE EXECUTED OR FOLLOWED WITH THE UT...

Страница 9: ...Operating system 4 GB of RAM 4 GB of available hard drive space VGA color monitor with minimum screen resolution 1024 x 768 A USB interface for TS Series and Model 3411 LAN can be also used for Model...

Страница 10: ...INTRODUCTION TORQUE 10 2 www magtrol com...

Страница 11: ...2020 NI DAQmx Runtime 20 1 NI Serial Runtime 20 0 NI VISA Runtime 20 0 Navigate inside the USB flash drive DVD Rom under TORQUE 10 folder Then Run install exe Figure 2 Run Setup Windows Use Explorer t...

Страница 12: ...ure 3 Destination Directory Window This process can take several minutes to an hour depending upon the computer being used Please plan accordingly When all programs are installed the following window...

Страница 13: ...r x86 msi 32 bit operating system Double click on the correspondent driver 3 2 MODEL 3411 DRIVER The MODEL 3411 can be controlled in USB TMC interface or COMx CDC interface TORQUE 10 supports the USB...

Страница 14: ...3411 LAN Settings 3 2 1 LAN CONFIGURATION Run NI MAX exe from National Instruments software tools Open Devices and Interfaces Right click on Network Devices select Create New VISA TCP IP Resource Sele...

Страница 15: ...ATION www magtrol com 7 Enter IP address and Port Number 3411 then press Validate Figure 6 NI MAX TCP IP Press Validate Figure 7 NI MAX Validate Press OK and Next Enter the name of port which will app...

Страница 16: ...some problem to configure correctly the LAN please ask your IT service to have some help 3 3 MODEL 3410 DRIVER No specific driver requested This MODEL 3410 has a true RS 232 interface RS 232 cable Fem...

Страница 17: ...10 TORQUE 10 EXECUTATION www magtrol com 9 4 TORQUE 10 EXECUTATION Double click on TORQUE 10 exe Figure 9 Main window 4 1 MENU On top of the window 3 menus are available 4 1 1 FILE MENU Figure 10 Fil...

Страница 18: ...windows opens to select the filename to save Save Data as Press to save the current data in specific filename The filename and path will be use when press Save Data Export to Excel Press to export dir...

Страница 19: ...TORQUE 10 SETUP CONFIGURATION www magtrol com 11 5 SETUP CONFIGURATION This is the default button when starting the TORQUE 10 program Figure 13 Setup display...

Страница 20: ...2BCF 0x0030 C 0052 2 INSTR Chapter TS 1xx Setup 3411 USB0 0x2BCF 0x0020 14L34110070 2 INSTR or LAN Chapter 3411 Setup 3410 COMx Chapter 3410 Setup To know the COMx port select go to Device Manager win...

Страница 21: ...can only be activated on one channel at a time Torque Unit Select the Units that you wish to display the Torque value Torque Filter Select the Torque Filter value Torque Invert Select the sign of Tor...

Страница 22: ...o acquire the Angle parameter HS Data Acquisition 1 Enable the High Speed acquisition 488 S sec T S Source 1 Select the source of data in High Speed mode Out Speed Scale Factor Set the transfer of ana...

Страница 23: ...CONFIGURATION www magtrol com 15 Figure 17 3411 USB setup 5 3 1 3411 LAN SETTING Figure 18 3411 LAN Setting Enter correct parameter for LAN settings Please contact your IT administration if you need m...

Страница 24: ...ication of the device Rated Torque Set the nominal torque of the Torquemaster connected to the unit Encoder Selects the pulse count per revolution of the encoder being used Torque Unit Selects the tor...

Страница 25: ...ble to activate two channels simultaneously All the instruments can be combined with each other in any combination It is important to note that the HS Data Acquisition mode cannot be activated when bo...

Страница 26: ...re information on TI Task refer to Section 10 1 Figure 23 Sensor Input Setting CONTROL FUNCTION Device Enable Check this box to enable the accessory temperature input device Channel 0 n The channel nu...

Страница 27: ...or INDICATOR FUNCTION Setup File Path Show the current Setup file path and name Data File Path Show the current Data file path and name Right click on field can be used to Open the file location This...

Страница 28: ...SETUP CONFIGURATION TORQUE 10 20 www magtrol com...

Страница 29: ...s will add parameters in Selected listbox The number of parameters in the list depends on the number of channels activated in the Setup page 6 2 SELECTED The selected parameters that will be measured...

Страница 30: ...DISPLAY CONFIGURATION TORQUE 10 22 www magtrol com...

Страница 31: ...ET Tare CH1 CH2 Set the current torque to the Tare value The tare is not saved SAVE Tare CH1 CH2 Set the current torque to the Tare value The tare is saved inside the TS RST Tare CH1 CH2 Remove the Ta...

Страница 32: ...nd if 1 or 2 channels are selected Vales for normal mode HS data acquisition disable The following table shows the Sampling rate per second MODEL Channel 1 or 2 Channel 1 2 High Speed COMx USB COMx US...

Страница 33: ...TEST NORMAL TWO CHANNELS Press RUN button to start the measurement Figure 29 Run Normal mode Two Channels This example shown a test at the maximum speed of acquisition in Normal mode 7 1 1 RESULT OF...

Страница 34: ...TEST FUNCTION TORQUE 10 26 www magtrol com 7 2 RUN TEST NORMAL SINGLE CHANNEL Figure 31 Run Normal mode Single Channels Figure 32 Acquisition from single channel...

Страница 35: ...data For the TS 100 Series this is the one shoot acquisition of 10 000 values of torque in 1 second For Model 3411 this is a continue acquisition at 488 Samples second Figure 33 HS configuration Figu...

Страница 36: ...Result of HS data acquisition Time stamp will be added for each torque value Figure 36 HS data acquisition Display Use File menu to save data Data are not automatically saved at the end of the test T...

Страница 37: ...10 TEST FUNCTION www magtrol com 29 7 4 EXAMPLE OF DATA FILE Figure 37 Data file example This file mdf is a text file formatted with tabular separator Use Torque 10 Notepad Excel or others to open thi...

Страница 38: ...TEST FUNCTION TORQUE 10 30 www magtrol com...

Страница 39: ...DISPLAY www magtrol com 31 8 DATA DISPLAY Figure 38 Data Button This view shows the measurement in spreadsheet windows Use the scrollbar to navigate at the end of the file Use File menu to save export...

Страница 40: ...rt the table directly to Excel sheet Figure 39 Export to Excel function This operation requires to have a licenced Microsoft Excel 2009 or later installed on the computer Figure 40 Excel view The Exce...

Страница 41: ...gure 41 Printer function dialog The printer cannot be selected for this operation the spreadsheet will be printed on the Default printer of Windows The default printer appears on the last line of dial...

Страница 42: ...DATA DISPLAY TORQUE 10 34 www magtrol com...

Страница 43: ...e plot with your mouse or moved with the left and right cursor buttons CREATE A CURSOR Right click in the cursor window Select Create Cursor and Single Plot Cursor 0 will now appear in the plot area a...

Страница 44: ...h and zoom in or out on small portions of the plot Comment You may type any comments you wish to retain with the data for this test and also print on the hard copy of the graph Use File menu to save e...

Страница 45: ...ature input task double click the Measurement Automation Explorer MAX shortcut on your desktop Figure 46 Expand Data Neighborhood by clicking on the small triangle next to it If you see NI DAQmx Tasks...

Страница 46: ...ww magtrol com Click NI DAQmx Task and then Next Figure 49 Skip to step 7 Right click NI DAQmx Tasks and click Create New NI DAQmx Task Figure 50 Expand Acquire Signals Figure 51 Expand Analog Input F...

Страница 47: ...ct being configured for thermocouple temperature measurement is a NI 9211 in the cDAQ chassis This is a four channel device and the tree will show the available channels Expand cDAQ1Mod1 NI 9211 Figur...

Страница 48: ...ation settings for this thermocouple module Figure 59 To configure all channels the same click the first channel scroll down to the last channel then press the Shift key and click on the last item Sel...

Страница 49: ...limited to thermocouple inputs you may select any of the temperature input options that the device is capable of providing You can change parameters for each individual channel of that device by selec...

Страница 50: ...CREATING TASK TORQUE 10 42 www magtrol com...

Страница 51: ...1 Visit Magtrol s Web site at https www magtrol com magtrol return material authorization rma us https www magtrol com magtrol return material authorization rma us to begin the RMA process 2 Complete...

Страница 52: ...n of returned goods Value for customs purposes only This value should not exceed 1000 CHF for repair units Origin of the goods Switzerland or USA Noticed failures or complaints 3 A cost estimate for s...

Страница 53: ...at any time on Magtrol s website www magtrol com in the SUPPORT section Please compare the date of this manual with the revision date on the web site then refer to the manual s table of revisions for...

Страница 54: ...al 2 edition 07 2023 MAGTROL INC 70 Gardenville Parkway Buffalo NY 14224 USA phone 1 716 668 55 55 fax 1 716 668 87 05 e mail magtrol magtrol com MAGTROL SA Rte de Montena 77 1728 Rossens Switzerland...

Отзывы: