Chapter 8 – Troubleshooting
MAINTENANCE
9. Select resistors until two adjacent resistance values, within 1% or 2% of each other, cause
the reading to swing through zero. Do not use any resistance below 10K.
10. Once the correct resistance value has been established, a high quality 50 ppm/°C, 1% or
better, precision resistor (RN60C or RN65C) must be obtained. Permanently solder it onto
the board in the R20 position.
11. The ZERO control potentiometer, which was previously set to a mechanical center, should
now trim out any remaining imbalance.
Note:
If something has caused the original balance to shift to such an
extent that the rear panel ZERO balancing control is out of range,
it is probable that more difficulties lie ahead. Whereas a new load
cell might offer the best solution, these instructions may get the
dynamometer operational.
8.1.2.2
Erratic Torque Reading
If the torque reading is erratic, then the problem could be a defective load cell or electronic component
in the dynamometer or readout device. It will be necessary to establish specifically what is defective.
At this time contact Magtrol Technical Assistance.
8.1.3
f
ull
S
Cale
T
orque
If the dynamometer cannot obtain full torque and will not support the beam and weight at full scale,
a resistance or current check may be used to help locate the problem.
8.1.3.1
Resistance Check
1. Disconnect the 2-pin connector and the dynamometer line cord.
2. Measure the resistance across the 2-pin male plug at the dynamometer.
3. Check the reading to the table below.
Dynamometer
Model
Resistance @ 20 °C
Ω
HD-106
171
HD-100
180
HD-400
80
HD-500
75
HD-510
75
HD-505
37.5
HD-515
37.5
HD-700
80
HD-710
80
HD-705
40
HD-715
40
HD-800
20
HD-810
20
HD-805
10
HD-815
10
HD-825
5
ED-715
40
ED-815
13