MAGSTIM 2002 Скачать руководство пользователя страница 3

Содержание 2002

Страница 1: ...September 2011 Magstim 2002 Operating Manual MOP01 EN Revision 01 CE 0086 BTL EX2029 Allergan v BTL PGR2021 00025...

Страница 2: ...peration 1O Section 5 Maintenance 11 5 1 User Maintenance and Calibration 11 5 2 Technical Maintenance 11 5 3 Voltage Selection and Fuse Rating 12 5 4 Cleaning and Disinfecting 12 5 5 Servicing 12 5 6...

Страница 3: ...t must be filed promptly with the transportation company All correspondence concerning the equipment must specify the model name and or number as well as the serial number exactly as they appear on th...

Страница 4: ...or users with cardiac demand pacemakers implanted defibrillators and or implanted neurostimulators Cortical magnetic stimulation runs the risk of inducing seizures Consideration should be given by the...

Страница 5: ...1 should be connected to this output Caution the system generates high intensity magnetic pulses through its coils which induce eddy currents in any conductive medium such as the human body nearby met...

Страница 6: ...an or to use it whilst pregnant should contact The Magstim Company Limited for more information Environmental Conditions The system must not be used or stored under environmental conditions that fall...

Страница 7: ...g functions described in this section assume that the mains power switch located on the rear panel is switched on 3 1 1 On Off Button 1 The on off button switch toggles the Magstim 2002 between on and...

Страница 8: ...mperature of the coil is below 5 C the system will not operate The Magstim 2002 will remain in its inactive default condition until it detects that the coil temperature has risen above 5 C 3 1 7 Repla...

Страница 9: ...Magstim 2002 can be disarmed by pushing the red stop button In this mode the disarmed indicator will illuminate and the instrument will discharge internally Note The Magstim 2002 will disarm if the c...

Страница 10: ...rigger the instrument when one of the trigger buttons located on the coil is also pressed The instrument will continuously charge and discharge if the trigger buttons are pressed and held 3 2 4 Coolin...

Страница 11: ...in Section 7 3 Please note The use of the correct accessories is essential to the functioning of the Magstim 2002 The Magstim Company Limited cannot guarantee the instrument s performance unless acce...

Страница 12: ...ed insert the pneumatic connector of the foot switch to the rear panel Foot Switch Socket on the Magstim 2002 unit 2 If an external triggering or recording device is to be used connect it to the Isola...

Страница 13: ...t the power level to the desired setting 4 4 Operation 1 Press the green run button to arm the Magstim 2002 Note If the system is not triggered within a minute of being armed it will revert back to th...

Страница 14: ...nector which causes damage to the contacts Continued use in this condition will eventually cause the contacts to deteriorate and loose electrical connection which will cause damage to the system Repai...

Страница 15: ...oltage Failure to do so may cause serious damage to the device s 5 4 Cleaning and Disinfecting The Magstim 2002 and accessories cannot be sterilised therefore do not allow them to become contaminated...

Страница 16: ...hould not be used in the vicinity of any equipment that does not comply with EMC Safety Standard EN 60601 1 2 including mobile phones Any interface cable which connects the Magstim 2002 unit to an ext...

Страница 17: ...Voltage Max Voltage Logic Low Voltage Min Voltage 2 x 10 6 discharges at 70 power level 8 x 10 5 discharges at 80 power level 4 x 10 5 discharges at 90 power level 2 x 10 5 discharges at 100 power lev...

Страница 18: ...being considerably heavier than the left Care should be taken when transporting the unit to ensure that this heavier side is adequately supported Figure 6 1 Guidelines on lifting and lowering 6 5 Pack...

Страница 19: ...il Website 44 0 1994 240798 44 0 1994 240061 info magstim com www magstim com 7 2 Servicing Enquiries Telephone Fax E mail 44 0 1994 242900 44 0 1994 240061 service magstim com 7 3 Sales Enquiries Tel...

Страница 20: ...rature sensor circuit failure C06 Bad Serial command or received data checksum CO Internal Software Error C08 Invalid Coil Temperature ID C09 Invalid temperature algorithm coefficient C10 Coil control...

Страница 21: ...max difference Stack over temperature Stack under temperature Stack max difference HVCAP over temperature HV Transformer over temperature Charge threshold failure VREF check failure HVCAP voltai ie c...

Страница 22: ...Aux 5V Pin 10 Trigger Edge Level When set to edge trigger the Magstim 2002 will only be triggered when the logic level being applied on the Tin or Tin pin is changed On the Tin pin triggering will oc...

Страница 23: ...turbances are controlled The customer or the user of the Magstim 200 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equip...

Страница 24: ...Meets requirement Mains power quality should be that of a typical commercial or interruptions and 95 dip in Ur for hospital environment If the user of the Magstim 200 2 requires voltage variations 0 5...

Страница 25: ...recommended separation distance in metres m Field Strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency rang...

Отзывы: