![Magnescale MG80-EI Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/magnescale/mg80-ei/mg80-ei_operating-manual_693840024.webp)
23
5.3.7.
Comparator
Up to 8 sets of comparetor settings are available for each frame.
Each comparator has option of 2 modes (2 or 4 steps of threshold).
Output will be the “area” number. Examples are shown below.
Example of comparator judgment
5.3.8.
Addition/Subtraction
This function will allow calculation (Addition/Subtraction) of the two different
measuring units and output result in a frame. The calculation result will be current
(Real) value. All 16 frames are available with this function. The calculation formula
is shown below.
Axis calculation result (current value)
=
<Sign1>Measurement value of measuring unit Ⓐ
+
<Sign2>Measurement value of measuring unit Ⓑ
(Example)
Measurement value of measuring unit Ⓐ
= 10[μm]
Measurement value of measuring unit Ⓑ
= 5[μm]
Calculation setting:Sign1 = +, Sign2 = -
Result after calculation(Current value) =
(+10) +
(-5)
= 5[μm]
Current value 12mm
<
Mode = 2 steps
>
Threshold 1 = 5mm
Threshold 2 = 20mm
Current value = 12mm
Output
→
a
rea “
1
”
<
Mode = 4 steps
>
Comparator threshold 1 = 5mm
Comparator threshold 2 = 10mm
Comparator threshold 3 = 15mm
Comparator threshold 4 = 20mm
Current value = 12mm
Output
→
area “2”
Threshold 1
Threshold 2
Current value 12mm
Threshold 1
Threshold 2
Threshold 3
Threshold 4