background image

 (E) 

1

1. Operating Cautions

Do not use the unit in places where it may receive excessive shocks.  Otherwise the inside of
the unit may be damaged or the unit may become unable to produce normal output signals.

Be sure to turn off the power before connecting or disconnecting connectors in order to
prevent damage or misoperation.

Insert the connector into the display unit until it locks. Be sure to turn off the power
switch before connecting or disconnecting the connector.

Locate the measuring unit at least 10 cm/3.94" away from a strong magnetic source.

Do not forcibly pull the cable for connecting or disconnecting, or it may cause breakage.

The measuring unit is short circuited to the frame GND (ground terminal) of the display
unit by shield wire of the cable.
When the measuring unit is used with an other signal processing device on a machine
tool, etc., be aware of the ground level.

Do not place the measuring unit where it is exposed to splash of water or oil.

In operating the feeler with a vacuum pump, use such an air-pass system as shown in
Fig. 1 to enable air driving. The optimum vacuum rate is 0.04 to 0.067 MPa. Further, put
such an orifice as shown in Fig. 2 on a tube from the air lifter connector to control the air
suction and discharge speed. The feeler is lifted at the air discharge to the vacuum pump.

Recommended calibration interval 1 year.

This unit is equipped with an anti-dust bellows on the spindle. Organic solvents, ozone,
ultraviolet rays, and other adverse conditions in the ambient atmosphere can cause rapid
deterioration of the anti-dust bellows. When using in these environments, either replace
anti-dust bellows periodically (every six months to one year), or use the organic solvent
resistant anti-dust bellows. (For ordering the organic solvent resistant anti-dust bellows,
please inquire at your Magnescale Co., Ltd. sales office or the Service Center.)

Unit: mm/inch

DK812AFLR / DK812AFLR5 /

DK812BFLR / DK812BFLR5

Fig.1 Air-pass System

Fig.2 Dimensions of Orifice

VG

EXH

Valve

To fitting tube

To vacuum
system

Fitting

Orifice

Filter

Cable

3/0.12"

 3/0.12"

3/0.12"

φ

0.3/0.01" dia.

φ

3/0.12" dia.

C0.5/0.02"

1/0.04"

Vacuum pump
system

φ

2.5/0.1" dia.

VAC

Содержание DK812AFR

Страница 1: ...itale Messtaster Read all the instructions in the manual carefully before use and strictly follow them Keep the manual for future references Lesen Sie die ganze Anleitung vor dem Betrieb aufmerksam du...

Страница 2: ...ICES 003 OPERATION IS SUBJECT TO THE FOLLOWING TWO CONDITIONS 1 THIS DEVICE MAY NOT CAUSE HARMFUL INTERFERENCE AND 2 THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED INCLUDING INTERFERENCE THAT MAY C...

Страница 3: ...J 1...

Страница 4: ...2 J...

Страница 5: ...J 3...

Страница 6: ...4 J...

Страница 7: ...J i 1 1 2 2 3 3 4 4 5 5 5 1 5 5 2 6 5 3 6 6 7 7 8...

Страница 8: ...ii J...

Страница 9: ...J 1 1 OFF 10 cm GND 1 0 04 0 067 MPa 2 1 6 1 DK812AFLR DK812AFLR5 DK812BFLR DK812BFLR5 1 2 mm VG VAC EXH 0 3 3 3 3 3 C0 5 2 5 1 VAC...

Страница 10: ...2 J 2 2 5 0 05 0 06 N m 9 7 0 15 mm DK812F 6 11 mm 0 02 mm 14 mm 1 0 6 1 0 N m 2 0 N m 40 mm DZ 191 DZ 191 DZ 191 2 M4 4 5 mm 2 1 mm O M4 4 5 3 9...

Страница 11: ...J 3 3 DK812AFLR DK812AFR DK812AFR5 DK812BFR DK812BFR5 BOX DK812AFLR 9 5 mm...

Страница 12: ...AFLR5 DK812BFLR5 A B Z EIA 422 40 m 0 1 m 0 5 m 2 mm 12 mm 1 mp p 1 5 mp p 0 6 0 3 N 0 5 0 3 N 0 4 0 3 N 1 1 mm 80 m min 42 m min 250 m min 100 m min DK812AFLR DK812AFLR5 DK812BFLR DK812BFLR5 100 m s2...

Страница 13: ...DK800B 80 m min 42 m min DK800A 50 ns A 1 200 ns 5 MHz DK800B 100 ns A 1 400 ns 2 5 MHz 5 1 5 1 DK800A 50 ns DK800B 100 ns 50 ns 100 ns A B DK800A 50 ns DK800B 100 ns A B Z A B 50 ns 200 ns 5 MHz A B...

Страница 14: ...00 ms Hi 5 3 5 V 5 CE22 A B 50 ns 100 ns 300 ns 500 ns A 1 200 ns 400 ns 1 2 s 2 s 5 MHz 2 5 MHz 833 kHz 500 kHz 0 1 m 0 5 m 80 m min 250 m min 42 m min 100 m min 14 m min 33 m min 8 4 m min 20 m min...

Страница 15: ...BFLR DK812BFLR5 CK T12 1 m CK T13 3 m CK T14 5 m CK T15 10 m 10 11 8 13 13 1 3 5 10 m 4 8 6 2 24 3 35 5 49 5 14 5 5 30 5 41 5 mm 5 5 13 6 13 6 55 5 1 3 5 10 m 41 5 13 6 50 CE22 01 1 m CE22 03 3 m CE22...

Страница 16: ...FR DK812AFR5 DK812BFR DK812BFR5 DK812AFLR DK812AFLR5 DK812BFLR DK812BFLR5 t 15 2 mm 12 29 7 5 8 M9 105 2 m 101 0 3 m 28 4 8 9 5 4 8 3 14 14 5 4 2 105 12 9 5 10 21 29 7 8 5 8 4 2 9 M9 97 mm t 0 3 8 110...

Страница 17: ...CS 1 Magnescale Co Ltd...

Страница 18: ...2 CS...

Страница 19: ...CS 3...

Страница 20: ...4 CS...

Страница 21: ...CS i 1 1 2 2 3 3 4 4 5 5 5 1 5 5 2 6 5 3 6 6 7 7 8...

Страница 22: ...ii CS...

Страница 23: ...CS 1 1 10 GND 1 0 04 0 067 MPa 2 1 6 1 Magnescale Co Ltd DK812AFLR DK812AFLR5 DK812BFLR DK812BFLR5 1 2 VG VAC EXH 0 3 3 3 3 3 C0 5 2 5 1 VAC...

Страница 24: ...2 CS 2 2 5 0 05 0 06 N m 9 7 0 15 DK812F 6 11 0 02 14 1 0 6 1 0 N m 2 0 N m 40 DZ 191 DZ 191 DZ 191 2 M4 4 5 2 1 O M4 4 5 3 9...

Страница 25: ...CS 3 3 DK812AFLR DK812AFR DK812AFR5 DK812BFR DK812BFR5 DK812AFLR 9 5...

Страница 26: ...K812AFLR DK812BFLR DK812AFLR5 DK812BFLR5 A B Z EIA 422 40 m 0 1 m 0 5 m 2 12 1 mp p 1 5 mp p 0 6 0 3 N 0 5 0 3 N 0 4 0 3 N 1 80 42 250 100 DK812AFLR DK812AFLR5 DK812BFLR DK812BFLR5 100 m s2 1000 m s2...

Страница 27: ...DK800A 80 DK800B 42 50 ns DK800A A 200 ns 5 MHz 100 ns DK800B A 400 ns 2 5 MHz 5 1 5 1 DK800A 50 ns DK800B 100 ns 50 ns 100 ns DK800A A B 50 ns DK800B 100 ns A B Z A B 50 ns 200 ns 5 MHz A B 100 ns 4...

Страница 28: ...ns DK800B 100 ns 5 2 A B 400 ms Hi 5 3 5V 5 CE22 A B 50 ns 100 ns 300 ns 500 ns A 200 ns 400 ns 1 2 s 2 s 5 MHz 2 5 MHz 833 kHz 500 kHz 0 1 m 0 5 m 80 250 42 100 14 33 8 4 20 DK800A DK800B A B Hi A B...

Страница 29: ...2AFLR5 DK812BFLR DK812BFLR5 CK T12 1 CK T13 3 CK T14 5 CK T15 10 10 11 8 13 13 1 3 5 10 4 8 6 2 24 3 35 5 49 5 14 5 5 30 5 41 5 5 5 13 6 13 6 55 5 1 3 5 10 41 5 13 6 50 CE22 01 1 CE22 03 3 CE22 05 5 C...

Страница 30: ...DK812AFR DK812AFR5 DK812BFR DK812BFR5 DK812AFLR DK812AFLR5 DK812BFLR DK812BFLR5 t 15 2 12 29 7 5 8 M9 105 2 101 0 3 28 4 8 9 5 4 8 3 14 14 5 4 2 105 12 9 5 10 21 29 7 8 5 8 4 2 9 M9 97 t 0 3 8 2 4 5 1...

Страница 31: ...e Safety Precautions before operating installing maintain ing inspecting repairing or otherwise working on this unit Warning Indication Meanings The following indications are used throughout this manu...

Страница 32: ...fire or electric shock Do not disassemble or modify the unit as this may result in injury or electric shock These actions may also damage the internal circuitry Caution Failing to follow the precautio...

Страница 33: ...properly Provide adequate safety measures to prevent damages in case our products should develop malfunctions Use outside indicated specifications or purposes and modification of our products will vo...

Страница 34: ...4 E...

Страница 35: ...Cautions 1 2 Mounting Instructions 2 3 Names of parts 3 4 Specifications 4 5 Measuring unit output signals 5 5 1 Output Signal Phase Difference 5 5 2 Output Signal Alarm 6 5 3 Receiver 6 6 Optional Ac...

Страница 36: ...ii E...

Страница 37: ...the feeler with a vacuum pump use such an air pass system as shown in Fig 1 to enable air driving The optimum vacuum rate is 0 04 to 0 067 MPa Further put such an orifice as shown in Fig 2 on a tube...

Страница 38: ...t Fig 1 Note The tightening torque should be roughly that needed to completely flatten the wave washer Recommended tightening torque 0 6 to 1 0 N m Maximum tightening torque 2 0 N m Fix the cable in a...

Страница 39: ...n DK812AFLR The air lifter cannot be attached to the DK812AFR DK812AFR5 DK812BFR DK812BFR5 Cable Interpolation box Connector DK812AFLR Feeler Air lifter connector Stem diameter 9 5 mm 0 37 Anti dust b...

Страница 40: ...2BFR5 DK812AFLR DK812BFLR DK812AFLR5 DK812BFLR5 A B Z phase voltage differential line driver output compliant with EIA 422 40 m 0 1 m 0 5 m 2 mm 0 08 12 mm 0 47 1 mp p 1 5 mp p Downward 0 6 0 3 N Hori...

Страница 41: ...than 50 ns for the DK800A phase A cycle 200 ns 5 MHz or smaller than 100 ns for the DK800B phase A cycle 400 ns 2 5 MHz The minimum phase difference can be modified under special specifications See 5...

Страница 42: ...he supply voltage is 5 V 5 Use the CE22 series extension cables for bare wires optional accessories A B minimum phase difference 50 ns 100 ns 300 ns 500 ns Phase A cycle 200 ns 400 ns 1 2 s 2 s Counte...

Страница 43: ...m 32 8 11 8 0 46 dia 13 0 51 Cable length 1 3 5 10 m 3 3 9 8 16 32 8 4 8 0 19 6 2 0 24 24 3 0 96 35 5 1 4 49 5 1 95 14 0 55 5 5 0 22 30 5 1 2 41 5 1 63 Unit mm inch 5 5 0 22 13 6 0 54 13 6 0 54 55 5...

Страница 44: ...12 0 47 dia 29 7 1 17 5 8 0 23 M9 105 4 13 Cable length 2 m 6 6 101 3 98 Cable length 0 3 m 0 98 8 0 31 4 8 0 19 dia 9 5 0 37 dia 4 8 0 19 dia Tightening nut Wave washer 3 0 12 14 0 55 dia 14 5 0 57...

Страница 45: ...erheitsma nahmen vor der Inbetriebnahme Installation Wartung Inspektion oder Reparatur dieses Ger tes oder der Durchf hrung anderer Arbeiten durch Bedeutung der Warnhinweise Bei der Durchsicht dieses...

Страница 46: ...rufen werden kann Das Ger t nicht zerlegen oder ver ndern da dies zu Verbrennungen oder elektrischen Schl gen f hren kann Durch derartige Ma nahmen k nnen auch die internen Stromkreise besch digt werd...

Страница 47: ...dass unsere Produkte einwandfrei funktionieren Sorgen Sie f r geeignete Sicherheitsma nahmen um im Falle von Ger test rungen Sch den auszuschlie en Wenn das Profukt modifiziert oder nicht seinem Zweck...

Страница 48: ...4 G...

Страница 49: ...besonderen Beachtung 1 2 Montageanleitung 2 3 Teilebezeichnungen 3 4 Technische Daten 4 5 Messtaster Ausgangssignale 5 5 1 Ausgangssignal Phasendifferenz 5 5 2 Ausgangssignalalarm 6 5 3 Empf nger 6 6...

Страница 50: ...ii G...

Страница 51: ...etzt ist F r den Betrieb des F hlers mit einer Unterdruckpumpe ist das Druckluft system entsprechend Abb 1 anzuordnen Der optimale Druck liegt bei 0 04 bis 0 067 MPa 0 4 bis 0 67 bar Dar ber hinaus is...

Страница 52: ...nweis Die Spannmutter ist so fest anzuziehen dass die Wellenscheibe vollkommen flachgedr ckt wird Empfohlenes Anzugsmoment 0 6 bis 1 0 N m Maximales Anzugsmoment 2 0 N m Das Kabel ist in einer geeigne...

Страница 53: ...AFLR Lediglich der Luftabheber ist am DK812AFR DK812AFR5 DK812BFR DK812BFR5 nicht anzubringen DK812AFLR F hler Luftabheberanschluss Schaft Durchmesser 9 5 mm Staubschutzfaltenbalg Spannmutter Wellensc...

Страница 54: ...12AFR DK812BFR DK812AFR5 DK812BFR5 DK812AFLR DK812BFLR DK812AFLR5 DK812BFLR5 A B Z Phasen Spannungsdifferential Leitungstreiberausgabe entspricht EIA 422 40 m 0 1 m 0 5 m 2 mm 12 mm 1 ms s 1 5 ms s Ab...

Страница 55: ...erger tes kleiner als 50 ns beim Modell DK800A A Phasen Zyklus 200 ns 5 MHz bzw 100 ns beim Modell DK800B A Phasen Zyklus 400 ns 2 5 MHz Die minimale Phasendifferenz kann unter Sonderspezifikationen m...

Страница 56: ...ll muss sichergestellt werden dass die Versorgungsspannung 5 V 5 ist Verwenden Sie Verl ngerungskabel der Serie CE22 f r die blanken Dr hte gesondert erh ltliches Zubeh r Minimale A B Phasen differenz...

Страница 57: ...BFLR5 verwenden Verl ngerungskabel CK T12 1 m CK T13 3 m CK T14 5 m CK T15 10 m 11 8 13 Kabell nge 1 3 5 10 m 4 8 6 2 24 3 35 5 49 5 14 5 5 30 5 41 5 Einheit mm 5 5 13 6 13 6 55 5 Kabell nge 1 3 5 10...

Страница 58: ...DK812AFR DK812AFR5 DK812BFR DK812BFR5 DK812AFLR DK812AFLR5 DK812BFLR DK812BFLR5 Dicke t 15 2 mm 12 29 7 5 8 M9 105 Kabell nge 2 m 101 Kabell nge 0 3 m 8 28 4 8 9 5 4 8 Spannmutter Wellenscheibe 3 14...

Страница 59: ...DK812AFR DK812AFR5 DK812AFLR DK812AFLR5 DK812BFR DK812BFR5 DK812BFLR DK812BFLR5...

Страница 60: ...DK812AFR DK812AFR5 DK812AFLR DK812AFLR5 DK812BFR DK812BFR5 DK812BFLR DK812BFLR5...

Страница 61: ...stin es exclusivement l usage des acqu reurs de l quipement d crit dans ce manuel Magnescale Co Ltd interdit formellement la copie de quelque partie que ce soit de ce manuel ou son emploi pour tout au...

Страница 62: ...DK812AFR DK812AFR5 DK812AFLR DK812AFLR5 DK812BFR DK812BFR5 DK812BFLR DK812BFLR5...

Страница 63: ...1 1 1 2 2 1 2 3 4 5 3 4 5 DK812AFR DK812AFR5 DK812AFLR DK812AFLR5 DK812BFR DK812BFR5 DK812BFLR DK812BFLR5...

Страница 64: ...in Japan 2007 Magnescale Co Ltd DK812AFR DK812AFR5 DK812AFLR DK812AFLR5 DK812BFR DK812BFR5 DK812BFLR DK812BFLR5 3 208 605 02 108 6018 15 18 Shinagawa Intercity Tower A 18F 2 15 1 Konan Minato ku Tokyo...

Отзывы: