MAGNAMED TECNOLOGIA MÉDICA S/A
6.5.4 ALARM
For each alarm directly related to the ventilation process, there are two limits (high value and
low value) to be adjusted. These limits are configured directly in ALARM menu.
To make these settings, touch the button for the alarm corresponding to alarm to be set, and
this one will be selected (the color will change), allowing the modification of its value using the spin and
confirm button.
It is also possible to change the limits value of the alarm through the slider (adjustment bar),
moving the finger by this control.
To confirm the set value, press the button again corresponding to the alarm or press the spin
and confirm button (ENTER).
In this menu, it is also possible to adjust the maximum time allowed of patient apnea, which will
determine the entry of backup ventilation.
There is also the option for auto setting of alarms and in order to enable, it is necessary that the
ventilator is not in STANDBY (Standby mode) and preferably the ventilation is stabilized, seeking greater
patient’s safety.
•
Whenever restarted the ventilator or the type of patient was changed, alarm limits shall
assume standard values, pursuant to the type of patient.
•
The apnea alarm time may be turned off. In this condition, BACKUP VENTILATION WILL
BE NEVER ACTIVATED. Therefore, the operator must be sure that this adjustment is
really necessary and aware of the clinical implications involved.
•
The automatic adjustment of alarm limits sets alarms for a percentage calculated on
the value monitored during ventilation, thus, it may only be adjusted when the
ventilator is NOT in standby mode (STANDBY) and preferably with an established
ventilator condition.
The alarm audio volume may also be changed. Just set the desired volume via the Alarms Audio
Volume button.