MD-2060 Series
© 2015 Magnaflux
Page 3
Magnaflux-SRVC-IG014
Issued: Apr 2015
Rev. 2
Model No.
Gallons
Liters
MD-2060 (2 door)
20
76
MD-2060L (3 door)
34
129
MD-2060XL (4 door)
46
174
MD-2060XL2 (5 door)
58
220
Recommended Bath Level:
When adding 14A particles to a new bath you can use
the following guide to achieve a 0.2 concentration. 1
level scoop of 14A particles per 1 gallon of Carrier II.
NOTE: DO NOT ADD ALL THE PARTILCES ALL AT
ONCE. This will cause particles to accumulate in the
bottom of the tank. Allow particles to mix for 20–30
minutes, then test concentration per instructions.
Duty Cycle:
At full output: 0.5 seconds “ON” – 10 seconds “OFF”
At 2000 amps: 0.5 seconds “ON” – 5 seconds “OFF”
The equipment can be fused based on less than the maximum amperage draw due to the duty
cycle. Refer to NEC Code Section 630.31 Ampacity of Supply Conductors and Table 630.31
(A) (2) Duty Cycle Multiplication Factors for Resistance Welders (NPFA National Electric Code
2011). Some areas refer to 2 phase instead of 1 phase. The incoming power goes directly to
an isolation transformer so the unit will operate identically on 1 or 2 phase power.
Electrical Input:
Input Volts AC
Phase
Full Load
(Amps)
Recommended
Line Fuse (Amps)
Recommended
Wire Size (AWG)
208
3
N/A
N/A
N/A
230
3
N/A
N/A
N/A
380
3
450
225
3/0
415
3
430
225
3/0
460
3
400
200
2/0
575
3
325
200
2/0
5. A compressed air supply capable of providing a minimum 90-120 psig at 5 cubic feet per
minute. Connect the incoming air supply to the Incoming Air Connection on the air filter ex-
ternally on the unit. Adjust the Air Pressure Regulator to 40 psig to supply air to the head-
stock extension cylinder. Follow all local codes for plumbing pneumatics.
6. Fill the tank to the appropriate level as per the Recommended Bath Level.
7. Connect the pump plug to the 120VAC receptacle that is located above the black box on
the headstock side of the unit. Never run the pump dry. The pump On/Off switch is located
on the Operator Control Panel.