CMD-2030 Series

 

© 2015 Magnaflux 

Page 3 

Magnaflux-SRVC-IG028 

Issued: Apr 2015 

Rev. 2 

The equipment can be fused based on less than the maximum amperage draw due to the duty 
cycle. Refer to NEC Code Section 630.31 Ampacity of Supply Conductors and Table 630.31
(A) (2) Duty Cycle Multiplication Factors for Resistance Welders (NPFA National Electric Code 
2011). Some areas refer to 2 phase instead of 1 phase. The incoming power goes directly to 
an isolation transformer so the unit will operate identically on 1 or 2 phase power.  

Electrical Input: 

Input Volts AC 

Phase 

Full Load 

(Amps) 

Recommended 

Line Fuse (Amps) 

Recommended 

Wire Size (AWG) 

208 

N/A 

N/A 

N/A 

N/A 

230 

240 

200 

#4 

380 

145 

125 

#4 

415 

130 

100 

#6 

460 

120 

100 

#6 

575 

100 

80 

#6 