Black Light Installation
(Optional Accessory)
1. Plug black light cord into the BLACK LIGHT outlet on either end of the inspection unit. Test ground fault
interrupt for proper operation.
2. If using the optional 600086 BLACK LIGHT with the integral mag shot button, plug the mag shot cable into
the REMOTE MAG socket conveniently located on either end next to the BLACK LIGHT outlets.
Adjustment of Magnetizing Shot Duration
The duration of the magnetizing shot is set at 0.5 seconds from the factory.
1. Figure 2-7 shows the PLC, while Figure 2-8 shows the port cover open on the PLC. Adjust the top small
potentiometer behind the communication port cover on the PLC. The range is from 0.5 to 5 seconds. Use
caution not to exceed the duty cycle of the unit when adjusting the timer. With the screw turned all the way
counter clockwise, that equals 0.5 seconds. If the screw is turned all the way clockwise, that equals 5
seconds. A shot timer (Magnaflux Accessory Part #622901) should be used to provide the actual shot time
that has been set.
Figure 2-7
Figure 2-8
Содержание CD-2000 Series
Страница 1: ...CD 2000 Series Revised January 2020 Page 1 0 CD 2000 SERIES...
Страница 7: ...CD 2000 Series Revised January 2020 Page 1 6 PAGE LEFT BLANK INTENTIONALLY...
Страница 14: ...CD 2000 Series Revised January 2020 Page 2 7 PAGE LEFT BLANK INTENTIONALLY...
Страница 19: ...CD 2000 Series Revised January 2020 Page 3 5 PAGE LEFT BLANK INTENTIONALLY...
Страница 24: ...CD 2000 Series Revised January 2020 Page 4 5 PAGE LEFT BLANK INTENTIONALLY...
Страница 30: ...CD 2000 Series Revised January 2020 Page 5 6 PAGE LEFT BLANK INTENTIONALLY...
Страница 33: ...CD 2000 Series Revised January 2020 Page 6 3 PAGE LEFT BLANK INTENTIONALLY...
Страница 34: ...CD 2000 Series Revised January 2020 Page 7 1 SECTION 7 HOOD INSTALLATION Not Applicable...
Страница 35: ...CD 2000 Series Revised January 2020 Page 7 2 PAGE LEFT BLANK INTENTIONALLY...