Technologist Control Unit Operation 

 
 
Terminology: TCU = Technologist Control Unit 
 
1: 

To turn Genesis MRI Music 

ON

 from TCU push the Green

 Power 

button (located on the 

top right corner). 

 
2: 

To adjust Technologist’s 

Speaker Music Level

 depress the Green 

TECH VOL+ 

and the 

Red 

VOL– 

(located to the left of the 

Power

 button) for UP and Down respectively. The 

Stereo will show the level.  To 

Mute

 the Technologist’s 

Speakers 

depress the Violet 

TECH 

MUTE

 button (located under the Red 

VOL-

 button). The “

ON

” will flash on the mid right 

side of the display in the 

MUTE

 box.   

 
3: 

On power up the microprocessor sets 

VOL, MIC, and AVC 

level

 

for patients wherever it 

was left when the system was turned off.  The number equals the percentage of volume 
and the horizontal bar gives a quick visual representation.  

 
4: 

Depressing the Violet 

MIC

 button (located on bottom left corner) will allow you to override 

music and talk to the patients.  Level can be adjusted buy depressing Green 

MIC VOL+, 

or 

the Red 

VOL-,

 for UP and Down respectively.  A good starting point is 50.  When 

MIC 

button 

is depressed you will see “MIC” on the lower left “

MODE

” box of the display. 

 
5: When 

the 

AVC (intercom integration)

 input has triggered from depressing the 

MIC

 button 

on your GE Intercom (providing the intercom integration kit was installed) the Display will 
show 

AVC 

in the lower left “MODE” box, and the music to the patient will be interrupted.  

The AVC level and the console intercom level will control the level to the patients (it’s a 
good idea adjust the AVC and, leave the console set the same for a Patient without 
Headsets.)  A good starting point is 50 but no two MRIs are the same. 

 
6: 

Depressing the Blue 

PAT OVERRIDE

 button (located above the 

MIC

 button) will toggle 

thru locking out the patient from adjusting the 

VOL, SEL, VOLSEL, or OFF

.  This will also 

be displayed on the lower right 

“PAT OVR”

 box. 

 
7: 

We recommend getting in the habit of depressing the Blue 

PAT MUTE

 button (located 

above 

PAT OVERRIDE

 button) after each patient (this way when you are setting up a new 

patient they will be able to hear you, if the music is on they probably can’t.) 

 
8: 

Depressing the Green 

VOL NORM 

button (located above 

VOL

 

OFF

 button) will raise 

Volume for the Patients to 20. This preset is a good starting point and can be adjusted in 
program mode described in 

step 14.

 

 
9: Depressing 

Green 

MUSIC

 

VOL-

 button (located above the 

VOL NORM 

button) will lower 

Volume for the Patients, and depressing the 

MUSIC

 

VOL+

 (located above the 

MUSIC

 

VOL-

   button) will raise Volume for the Patients. 

 
10:  The Genesis has a feature called

 AUTO GAIN 

this provides the ability to automatically 

adjust volume level of music to Patients dependent on the amount gradient noise being 
created by MRI.  

 

 

Содержание Genesis ULTRA

Страница 1: ...llation Manual Index System Notes 2 Headset Operation 3 Technologist Control Unit Operation 4 5 and 6 Important Safety Instructions 7 Pre Installation 8 System Parts List 9 10 Genesis ULTRA Installati...

Страница 2: ...hanges volume automatically to mask MR gradient sounds Patient volume and music selection controls with voice feedback for maximum comfort and versatility Backlit Technologist Control Unit allows oper...

Страница 3: ...to provide the highest level of attenuation it is necessary to first rotate tops of Headset toward the rear of the patients head then while applying pressure rotate fully forward to align with the ear...

Страница 4: ...roviding the intercom integration kit was installed the Display will show AVC in the lower left MODE box and the music to the patient will be interrupted The AVC level and the console intercom level w...

Страница 5: ...C VOL button will put the system in Program Mode Depressing the MUSIC VOL and VOL button toggle up and down through the Menu respectively Depressing the Blue and Violet INPUT ENTER button located bell...

Страница 6: ...6 Magnacoustics Inc 3 2008 rev o...

Страница 7: ...g periods of time 14 Refer all servicing to qualified service personnel Servicing is required when the apparatus has been damaged in any way such as power supply cord or plug is damaged liquid has bee...

Страница 8: ...ck for the following items 1 AM FM Multi Disc CD Changer 2 One AM antenna 3 One FM antenna 4 Two Speakers 5 One Transducer Module houses connections for Headsets Patient Volume control Patient selecti...

Страница 9: ...s over IR receiver on the Stereo 18 CB8 6ft AC Adapter Cable connects the AC Outlet to the Switched Voltage Selectable and Fused AC Receptacle on the rear of Genesis module 19 AutoVoice Breakout Box i...

Страница 10: ...minals on the back of the Stereo Red and Black to Left Red and Black Green and White to Right Black and Red respectively The other side of cable CB6 to the upper terminals of SPK1 on the rear of Genes...

Страница 11: ...Cable install a 75 to 300 ohm adapter to eliminate any loops and only connect 1 wire 16 For GE MRI s remove block off plate from J53 or J52 at penetration panel all others use supplied kit or fabricat...

Страница 12: ...rcuit B Power If not pull AC Power and contact Magnacoustics Tech Support at 1 800 637 2282 or E mail at support magnacoustics com for further details 36 Press the green Power Button on the TCU a self...

Страница 13: ...8 Tech Unit Connector J2 fMRI Input J9 RS232 Spare J3 Video Input J10 RS 232 Spare J4 MIC Input J11 To Transducer CB1 J5 AVC Input J12 AC Receptacle 5 amp SB J6 fMRI trigger output SPK1 Speaker Juncti...

Страница 14: ...eo Multi CD Player SCAN ROOM MRI Pen Panel Pneumatic Transducer Patient Headset Pneumatic Tubing IR Cable CB 5 CB 2 50 Technologist Control Unit Patient Mic Patient Controls DB9 Filter Gradient Mic Sp...

Отзывы: