7
Preparing for testing
In order for you to save time for both you and the client, we recommend that you prepare the environment, the client,
the test device setup and the OTOsuite software so that they are ready for the test.
To prepare for testing:
•
If this is the first test of the day, make a probe check. See
•
If needed, connect the contralateral phone. See
.
•
Prepare MADSEN Zodiac and the Immittance module: select the patient and test setup. See the OTOsuite User Guide.
•
Prepare the patient. See
.
•
Prepare probe and eartip. See
Fitting the eartip on the probe
and
Fitting the probe in the patient’s ear
7.1
Preparing the test environment
Immittance tests require no specific test environment with regard to noise.
7.2
Hygienic precautions
•
Be sure to follow any established infection control procedures for the setting in which you are working.
•
Always use clean eartips.
•
Swab the probe tip with disinfectant between ears. If one ear is infected, test the healthy ear first and change eartips
between ears.
•
To prevent cross-infection, use a clean eartip when you test the next patient.
7.3
Probe check
To make sure that the probe is functioning correctly, it is recommended that you perform a probe check at the start of
each day.
Caution
•
Always clean and disinfect the probe tip before you insert it into a test cavity.
See
Cleaning the probe and probe tip
Note
•
If the test environment changes, for instance if there is an increase in humidity or if you are going to test
at a different altitude, make a probe check to verify that the system measures correctly.
Note
•
If the probe check result does not show a value of 1.9 - 2.1 mmho/cc/ml at 226 Hz, we recommend that
you make an admittance calibration. See
36
Otometrics - MADSEN Zodiac
7 Preparing for testing
Содержание zodiac
Страница 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Страница 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Страница 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Страница 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Страница 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...