background image

Product Notes

Reading Intervals

The UltraShock measures pressure, temperature, humidity and tri-axial shock. When the 
device is running, the shock parameter samples at 512Hz and at the user specified reading 
rate, the device writes the peak values (g-force) from the samples to memory. 
For Example:

Pressure, temperature and humidity readings are taken at the user specified reading rate 
up to a maximum of 2 seconds intervals.

Axis Orientation

When the UltraShock is laying flat, with the label facing up, the 
x-axis indicates left to right direction and will read ~0g when 
still. The y-axis reads in a top to bottom direction and reads ~0g 
when still. The z-axis reads perpendicular to the other two axis, 
and will read ~1g when still.

Device Maintenance

Battery Replacement

Materials: 

3/32” HEX Driver (Allen Key) and a Replacement Battery

 

(U9VL-J)

- Remove the cover from the device by unscrewing the four screws.
- Remove the battery from its compartment and unsnap it from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw the enclosure back together 

securely.

 

Note: Be sure not to over tighten the screws or strip the threads.

Recalibration

The UltraShock standard calibration is performed at one temperature point at 25°C, two 
humidity points at 25%RH and 75%RH, two pressure points at 15PSI and 27-30PSI and 0g 
for the x-axis, 0g for the y-axis and 1g for the z-axis.

 

Note: The shock parameters are not traceable to NIST.

Pricing:

Recalibration traceable to NIST 

  $156.00

Recalibration 

 $136.00

Additional Services:

As Found Data 

  $15.00 per channel 

Verification Point 

  $15.00 per point 

Prices and specifications subject to change. See MadgeTech’s terms and conditions at www.madgetech.com

To send the devices back, visit www.madgetech.com, select 

Services

 then 

RMA Process

UltraShock

Product Quick Reference Card

Installation Guide

Installing the Interface cable

- IFC200
  Insert the device into a USB port. The drivers will install automatically. 

Installing the software

Insert the Software CD in the CD-ROM Drive. If the autorun does not appear, locate the 
drive on the computer and double click on 

Autorun.exe

. Follow the instructions provided 

in the Wizard.

Connecting the data logger

- Once the software is installed and running, plug the interface cable into the data logger. 
- Click the 

Communication Menu

, then 

Auto Configure Port

.

- After a moment, a box will appear stating a device has been found. 
- Click 

OK

. The 

Device Status

 box will appear. Click 

OK

.

- At this point, communications have been configured for your logger. These settings can 

be found under the 

Communication Menu

.

Note: For additional installation instructions refer to your “

Data Logger & Software Operating Manual”.

Device Operation

Starting the data logger

- Click 

Device Menu

 then

 Start Device

- Choose the desired start method.
- Choose the start parameters by selecting a 

Reading Rate

 suitable for your application.

- Enter in any other desired parameters and click 

Start

- A box will appear stating the data logger has been started. Click 

OK

- Disconnect the data logger from the interface cable and place it in the environment to 

measure.

 
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the 
device cannot be restarted until it has been re-armed by the computer.

Downloading

 

data

 

from a

 

data

 

logger

- Connect the data logger to the interface cable.
- Click the 

Device Menu

 then 

Read Device Data

. This will offload all recorded data onto 

the PC. 

Sampled Data:

Recorded Data:

Y-Axis

Z-Axis

X-Axis

Отзывы: