Wiring the Data Logger

Wiring Options

Three-position removable screw terminal 
connections; accepts 3-wire configurations. 

Warning: Note the polarity instructions. Do not attach wires to the wrong terminals.

 

Product Notes

LEDs

- Green LED blinks: 10 seconds to indicate logging and 15 seconds to indicate  

delay start mode.

- Red LED blinks: 10 seconds to indicate low battery and/or memory and 1 second  

to indicate an alarm condition.

Password Protection

An optional password may be programmed into the device to restrict access to 
configuration options. Data may be read out without the password

Multiple Start/Stop Mode Activation

- To start device: Press and hold the pushbutton for 5 seconds, the green LED will flash 

during this time. The device has started logging.

- To stop the device: Press and hold the pushbutton for 5 seconds, the red LED will flash 

during this time. The device has stopped logging.

Alarm Settings

Programmable high and low limits; alarm is activated when current reaches or exceeds set 
limits.

Engineering Units

Engineering units are used to convert one measurement reading to another. The 
MadgeTech software allows for software level Engineering Units (conversion applied to 
data after download). Certain devices have device level Engineering Units, which upon 
download automatically appear in the chosen unit of measure.

Please refer to the application note “Engineering Units”, found on the MadgeTech website, 
for information on how to manage Engineering Units. Also view the Engineering Units 
Video for step-by-step setup instructions. 

Installation Guide

Installing the Interface cable

- IFC200 

Insert the device into a USB port. The drivers will install automatically. 

Installing the software

The Software can be downloaded from the MadgeTech website at the following link: www.
madgetech.com/software-download. Follow the instructions provided in the Installation 
Wizard. 

 

Product User Guide

Process101A

Device Operation

Connecting and Starting the data logger

- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer. 
- The device will appear in the Connected Devices list, highlight the desired data logger.  
- For most applications, select “

Custom Start

” from the menu bar and choose the desired 

start method, reading rate and other parameters appropriate for the data logging 
application and click “

Start

”. (

Quick Start

” applies the most recent custom start options, 

Batch Start

” is used for managing multiple loggers at once, “

Real Time Start

” stores the 

dataset as it records while connected to the logger.

)

- The status of the device will change to “

Running

”, “

Waiting to Start

” or “

Waiting to 

Manual Start

”, depending upon your start method. 

- Disconnect the data logger from the interface cable and place it in the environment to 

measure.

 

Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this 

point the device cannot be restarted until it has been re-armed by the computer.

 

Downloading data from a data logger

- Highlight the data logger in the Connected Devices list. Click “

Stop

” on the menu bar.

- Once the data logger is stopped, with the logger highlighted, click “

Download

”. You will 

be prompted to name your report.

- Downloading will offload and save all the recorded data to the PC. 

Device Maintenance

Battery Replacement

Materials:

 

Small Phillips Head Screwdriver & Replacement Battery (LTC-7PN)

- Puncture the center of the back label with the screw driver and unscrew the enclosure.
- Remove the battery by pulling it perpendicular to the circuit board.
- Insert the new battery into the terminals and verify it is secure.
- Screw the enclosure back together securely.

Note: Be sure not to over tighten the screws or strip the threads.

Recalibration

 

The Process101A standard 
calibration is at two points. 
The points are dependent on 
the range of the data logger.

Additional Services:

Custom calibration and verification point options available, please call for pricing.

 
Call for custom calibration options to accommodate specific application needs.
Prices and specifications subject to change. See MadgeTech’s terms and conditions at 

www.madgetech.com

To send devices to MadgeTech for calibration, service or repair, please use the MadgeTech RMA Process by visiting  

www.madgetech.com

, then under the services tab, select RMA Process.

+ Input

- Input

Ground

Connector for interface cable

Range

20 mA

+160 mA

+3 A

Calibration Points

0 mA and 25 mA

0 mA and 100 mA

0 A and 1 A