Wiring the Data Logger

QuadVolt and OctVolt Single Ended Wiring (2.5V, 15V and 30V)

Two-position removable screw terminal connections; accepts 2-wire configurations.

QuadVolt and OctVolt Differential Wiring (+100mV)

Three-position removable screw terminal connections; accepts 3-wire configurations.

Warning: Note the polarity instructions. Do not attach wires to the wrong terminals. 

Product Notes

Engineering Units

Engineering units are used to convert one measurement reading to another. The 
MadgeTech software allows for software level Engineering Units (conversion applied to 
data after download). Certain devices have device level Engineering Units, which upon 
download automatically appear in the chosen unit of measure.

Please refer to the application note “Engineering Units” and Technical Support Videos, 
found on the MadgeTech website, for information on how to manage Engineering Units. 
Also view the Engineering Units Video for step-by-step setup instructions. 

Installation Guide

Installing the Interface Cable

- IFC200
  Insert the device into a USB port. The drivers will install automatically.

Installing the software

The Software can be downloaded from the MadgeTech website at the following link: 
www.madgetech.com/software-download. Follow the instructions provided in the 
Installation Wizard. 

Device Operation

Connecting and Starting the data logger

- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer. 
- The device will appear in the Connected Devices list, highlight the desired data logger.  

 

Product  User Guide

QuadVolt and OctVolt

- For most applications, select “Custom Start” from the menu bar and choose the desired 

start method, reading rate and other parameters appropriate for the data logging 
application and click “Start”. (“Quick Start” applies the most recent custom start options, 
“Batch Start” is used for managing multiple loggers at once, “Real Time Start” stores the 
dataset as it records while connected to the logger.)

- The status of the device will change to “Running”, “Waiting to Start” or “Waiting to 

Manual Start”, depending upon your start method. 

- Disconnect the data logger from the interface cable and place it in the environment to 

measure.

Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device 
cannot be restarted until it has been re-armed by the computer. 

Downloading data from a data logger

- Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “Download”. You will 

be prompted to name your report.

- Downloading will offload and save all the recorded data to the PC. 

Device Maintenance

Battery Replacement

Materials: 

3/32” HEX Driver (Allen Key) and Replacement Battery (U9VL-J)

- Remove the cover from the device by unscrewing the four screws.
- Remove the battery from its compartment and unsnap it from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw the enclosure back together 

securely.

Note: Be sure not to over tighten the screws or strip the threads.

Recalibration

 

The QuadVolt or OctVolt standard calibration is dependant upon the range. 

Additional Services:
Custom calibration and verification point options available, please call for pricing.

 

Connector for interface cable

+ Input

- Input

Range

100 mV

2.5 V

15 V

30 V

Calibration Point

0 mV and

90-100 mV

0 mV and

2.25-2.5 V

0 mV and

14.9-15.5 V

0 mV and

27-30 V

www.

.com

[email protected]

1.800.561.8187