MadgeTech DHS Скачать руководство пользователя страница 7

Ambient 

Temperature

Maximum Exposure 

Time (air)

-200 °C

128 minutes

-180 °C

137 minutes

-160 °C

148 minutes

-140 °C

163 minutes

-120 °C

183 minutes

-100 °C

213 minutes

-80 °C

263 minutes

-60 °C

368 minutes

-40 °C to +140 °C

Indefinitely

150 °C

601 minutes

160 °C

468 minutes

170 °C

396 minutes

180 °C

348 minutes

190 °C

313 minutes

200 °C

286 minutes

210 °C

265 minutes

220 °C

247 minutes

230 °C

233 minutes

240 °C

220 minutes

250 °C

209 minutes

260 °C

200 minutes

270 °C

192 minutes

280 °C

184 minutes

290 °C

178 minutes

300 °C

172 minutes

310 °C

166 minutes

320 °C

161 minutes

330 °C

157 minutes

340 °C

153 minutes

350 °C - 400 °C

60 minutes

Product User Guide

Time vs. Temperature Chart

www.

.com

[email protected]

1.800.561.8187

Содержание DHS

Страница 1: ...DHS Data Logging System DHS Data Logging System Dry Heat Sterilization Data Logging System Product User Guide www com information itm com 1 800 561 8187...

Страница 2: ...ta logger For most applications select Custom Start from the menu bar and choose the desired start method reading rate and other parameters appropriate for the data logging application and click Start...

Страница 3: ...lid of the thermal barrier 3 Unscrew the HiTemp140 M12 data logger from the lid of the thermal barrier 4 Slide the probe of the data logger out from the lid of the thermal barrier 2 Slide out the dat...

Страница 4: ...bsite 1 Visit the Services tab select RMA Process 2 When the web page opens please sign in If this is the rst time select Click here to register an account and create one Once signed in click on the M...

Страница 5: ...mmed into the device to restrict access to con guration options Data may be read out without the password Memory Wrap Around Yes Battery Type 3 6V high temperature lithium battery included user replac...

Страница 6: ...er to the Time vs Temperature Chart IP Rating Not Rated Barrier Materials 316 Stainless Steel PTFE Silicone Enclosure Dimensions 1 75 in OD x 9 6 in L 44 45 mm OD x 243 8 mm L Weight 3 0 lb 1350 g The...

Страница 7: ...minutes 170 C 396 minutes 180 C 348 minutes 190 C 313 minutes 200 C 286 minutes 210 C 265 minutes 220 C 247 minutes 230 C 233 minutes 240 C 220 minutes 250 C 209 minutes 260 C 200 minutes 270 C 192 mi...

Страница 8: ...ven if the exterior of the ThermoVault Max is cool to the touch the interior of the barrier and its contents may still be VERY hot The ThermoVault Max is primarily intended for use in dry air environm...

Отзывы: