Test the Instantaneous flow , UNIT:L/H
The K of MD-M12 K=17ml/P
1. Setting Procedure
1) Wiring; (Refer the Wiring)
2) Power on;
3) Press the
SET
key over 5s ,enter
to the setup mode
4)
Tap press
SET
key, rolling down setting the parameters;
5)
To set the
[C-D-P]
Modify the Decimal
places.To
press
the
lefe
key,the
parameters
is
flashing ,press the up or down key to adjust the place of
Decimal point.Press the set key to comfirm.
6)
To press the set key to set
[C-P] set the
total flow parameters . press the left key the number
flashing, press the up or down to set the parameter.
If the MD-M12 K=0.17Ml /p, to set the 0.017
(Unit: L)
7)
Press to set the
[F-P]
To set the instance
flow.If the MD-M12 K=17mL/p,The unit is ml the parameters