Troubleshooting Guide
Innowalk 2.gen
Document code:
T36
Made by:
FGB
Date:
25.03.21
Revision:
A
Approved by:
AL/KU
Date
:
25.03.21
Page:
2 of 12 pages
ELECTRONICS SIDE PANEL
Topic
No.
Parts to check
Actions
Still not fixed?
Product will not
power up
1
Check main power ON/OFF
switch
Turn the switch ON
Continue
troubleshooting
steps 2-5.
2
Check emergency stop button
Rotate the red mushroom button in the direction of the arrows to release
the emergency stop button.
Continue
troubleshooting
steps 3-5.
If faulty emergency
stop button.
Contact service
personnel to
replace button.
3
Check power cable
Visually inspect the power cable for damage. Unplug and insert the power
cable firmly into the machine and into the AC power socket.
Continue
troubleshooting
steps 4-5.
If broken or
damaged power
cable and/or
socket on machine.
Replace cable
and/or power
socket.
4
Check fuse
Unscrew the fuse holder and visually inspect the fuse. If broken, insert new
fuse according to specification printed on sticker on the switching panel.
(If the Innowalk works after changing the fuse, and there has been no
external power influences towards the Innowalk (eg.thunderstorms) please
contact your Made for Movement representative for a control of the
product)
Continue
troubleshooting
step 5.
5
Check power connection
Try different AC power outlet socket
Contact service
personnel