6 Error codes
The error code is displayed in hexadecimal
format.
The instrument identifies 4 errors which
may also occur simultaneously.
Hexadecimal format
Description
0080
Out Of Operating
Temperature
0400
Low Battery Voltage
1000
Metrological
verification period
expired
4000
Metrological Wrong
Checksum
Example of simultaneous errors
Low battery v Metrological verifica-
tion period expired: Err 1400
7 Test mode
ElecTo SJ can be temporarily set to a high
resolution in order to carry out measure-
ment tests.
Contact the manufacturer for further infor-
mation.