background image

MACURCO GAS DETECTORS 

Nitrogen Dioxide Detector Field Calibration Kit 

 ND-2 

 

CONTENTS OF THE KIT: 

1.

 

Gas Cylinder of 5 ppm Nitrogen Dioxide (NO

2

2.

 

Gas regulator with about two feet of plastic tubing 

3.

 

Calibration hood 

4.

 

These instructions 

 

GENERAL INFORMATION: 

All the Macurco Nitrogen Dioxide detectors are factory calibrated when they are shipped to the customer.  This 
calibration should be within specifications for 6 months to a year.  Some jurisdictions, local standards, or customer 
preferences require periodic recalibrations.  This Kit provides the means to fit those requirements.  Macurco 
recommends checking the operation of the the Macurco Nitrogen Dioxide detector every six months, either through use 
of the procedures given in the operating instructions or this Field Calibration Kit. 
 

ND-2 FIELD CALIBRATION PROCEDURE: 
NOTE:

  The ND-2 must have been power continuously for a minimum of 72 hours before proceeding 

 

Calibration Equipment Needed 

1.0

 

5ppm canister of Nitrogen Dioxide (NO

2

2.0

 

Digital multimeter, and set to DC volts, adjust for best resolution range (i.e. 2V) 

3.0

 

Second meter set to milliamps to 20mA range 

 

Preparation 

1.

 

Open the access door of the ND-2 case 

2.

 

Remove the two bottom screws that secure the interior faceplate, and swing the faceplate up. 

3.

 

Ensure that the digital display is "ON" before starting. 

4.

 

Turn "FAN" adjustment fully clockwise before proceeding. 

 

Calibration 

1.0

 

Connect meter leads across TP (Red lead + connects to the looped lead, nearest to display) 

2.0

 

Adjust RA5 Zero mult-turn 25K

Ω

 pot, until voltmeter reading is close to 0.00V (+/-0.06V) and adjust the 

milliamp output to 4.0mA +/- .03 via RA3 

 

 

 

 

Rev:  08/01 

 

Отзывы: