Macurco CD-xx H/MC Manual
REV
–
1.1
[34-2900-0510-8]
22
|
P a g e
4.5.12
4-20mA Output setting
–
“420”
To select the 4-20mA Output Option, in normal mode, push the
Next
button to get to “Con”
or the Configuration menu. Then
push the
Enter
button to enter the Con menu. The twelfth selection is the “420” or 4
-20mA Output Option. Push
Next
eleven
times to get to “420” then
Enter
. If the 4-
20mA is “On” push
Next
to turn it to “OFF” (flashing) then
push
Enter
to confirm the
change (solid) and push
Enter
again to return to “420” in the Con menu. Push
Next
until “End” is displayed then push
Enter
to
get back to normal operation.
5
Troubleshooting
5.1
On-Board Diagnostics
The CD-xxH/MC monitors all critical functions of the unit through software diagnostics that continuously test and verify unit
operations. If a problem is found, the unit will switch to a fail-safe/error mode or trouble condition. In this error mode, the
Alarm relay will be activated, the 4-20 mA current loop will go to 24 mA, the unit will display the error code, the green status
indicator LED light will flash, and the buzzer will chirp intermittently. The Fan relay will also engage if the Trouble Fan Setting
Option is set to “ON”. This is a s
afety precaution. To clear this mode, simply turn off power to the unit for a few seconds or push
the ENTER/TEST switch (inside the unit). This will cause the unit to restart the 1-minute self-test cycle.
5.1.1
4-20mA troubleshooting
•
0 mA is most likely a connection problem
•
4-20 mA is normal gas reading range (0-5000 ppm)
•
24 mA indicates a Trouble condition
5.1.2
“t” Error Codes
The CD-xxH/MC will display trouble codes
indicated by alternating every second “t
XX
” and “tYYY”
.
tXX
t01
Sensor Fatal Error
t02
Sensor Offset Regulation Error
t04
Sensor Algorithm Error
t08
Sensor Output Error
t10
Sensor Self Diagnostic Error
t20
Sensor Out of Range Error
t40
Sensor Memory Error
tYYY
t001
Missing Sensor (At Power Up only)
t002
At each power-up it checks if that ABC is disabled. If it is not disabled it will try to set it to disable. If it
fails it will trigger trouble t002.
t004
EEPROM bad checksum.
t008
Modbus communication error (during normal operation).
t010
Bad EEPROM
t020
Bad Factory calibration.
t040
Never Factory calibrated.
t080
Bad pressure during factory calibration
t100
Under range. Reading is under -100 ppm for more than 15 seconds
Содержание CD Series
Страница 31: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 31 P a g e 9 Appendix B Menu Structure 9 1 Main Menu ...
Страница 32: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 32 P a g e 9 2 Auto Test Menu bUZ ...
Страница 33: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 33 P a g e 9 3 Configuration Menu CON ...
Страница 34: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 34 P a g e ...
Страница 35: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 35 P a g e ...
Страница 36: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 36 P a g e ...
Страница 37: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 37 P a g e ...
Страница 38: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 38 P a g e ...
Страница 39: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 39 P a g e 9 4 Select Test Menu tst ...
Страница 40: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 40 P a g e 9 5 CAL Menu CD xxMC Only ...