PM100 Bump Test and Calibration Station 

REV – 1.0 

[34-2900-0513-2 ] 

15 

|

 

P a g e

 

 

The “Concentration” option will run the bump test until the test gas concentration is verified. If the units are in 

Stealth Mode, the Concentration option will be required for a successful bump test. 

The “Beep + Concentration” option will run the bump test until the audible alarm is detected and the test gas 

concentration is verified. 

5.3.1.3

 

Bump-Calibration Station Firmware Upgrade 

 

To upgrade the Bump-Calibration Station firmware,  

1.

 

Click 

Browse

.  

2.

 

Navigate to the location of the firmware file.  

3.

 

Click 

Open

.  

4.

 

The software interface will show the file path of the new firmware in the Station Firmware Upgrade field. 

Click 

Write

 to write the new firmware to the Bump-Calibration Station. 

5.3.2

 

PM100 Tab 

 

The Bump-Calibration Station and Bump-Calibration Station Manager software can be used to quickly make 

configuration changes for multiple PM

R

100

R

 detectors. The PM100 tab in the Bump-Calibration Station Manager 

Содержание AimSafety PM 100

Страница 1: ...IMPORTANT Keep these user instructions for reference AimSafety PMR 100R Bump Test Calibration Station Bump Test and Calibration Station for PMR 100 User Instructions ...

Страница 2: ......

Страница 3: ... Test Troubleshooting 8 4 3 1 Unit LED does not turn on or test is not performed 8 4 3 2 All tests fail 8 4 3 3 Test results not stored 8 5 Station Manager Software 8 5 1 Download and Installation 9 5 1 1 Download the software 9 5 1 2 Install the software 9 5 2 Connect to the Calibration Station 11 5 2 1 Connect Using PM Link 11 5 2 2 Connect Using USB Disk 12 5 3 Use the Software 12 5 3 1 Station...

Страница 4: ...t 1 844 325 3050 Do not disassemble unit or attempt to repair or modify any component of this instrument This instrument contains no user serviceable parts and substitution of components may impair product performance Using a certified gas with a concentration other than the one listed for this equipment when conducting a calibration verification test bump test will produce inaccurate readings Thi...

Страница 5: ...e IR communication of the PMR 100R Monitors to perform testing and calibration by pressing only a single button Operation of the station only requires the use of two buttons Learning to use the station takes only minutes It is designed to quickly perform bump tests calibration for multiple PMR 100R monitors It can test calibrate up to four detectors simultaneously which reduces the testing time an...

Страница 6: ... of this equipment by untrained or unqualified persons or use that is not in accordance with this user manual may adversely affect product performance 2 3 Specifications Size 18 50 x 16 14 x 8 66 47 x 41 x 22cm Weight 18 Lbs 8 2kg Without cylinder Operating Temperature 41 to 104 F 5 to 40 C Warranty 2 years Battery Type Rechargeable Lithium Ion Battery life 1000 Bump Tests LEDs 5 red green orange ...

Страница 7: ...scription Operation Status Unit LEDs On Testing On Test Failed On Test Passed Blinking 5 sec Communicating Power LED On Ready Blinking 5 sec No USB USB Full On Low Battery 3 3 Charge Battery The charging port is located on the left side of the unit The unit comes with a DC adapter which can be used to charge the unit using a 110V power supply A full charge will take approximately 3 hours and will ...

Страница 8: ...he USB memory NOTE All empty bays must have a rubber cap installed before a test can be performed Without the rubber cap installed excess test gas will leak out of the empty bay 4 1 Bump Test All monitors need to be periodically tested with a known concentration of gas to verify the sensor is responding correctly to the target gas This test is called a bump test To perform a bump test 1 Place the ...

Страница 9: ...t LED does not turn on or test is not performed Clean monitor IR window IR interference bright lights sunlight Monitor in gas alarm state 4 3 2 All tests fail Gas cylinder empty Check pressure gauge Gas pressure should read at least 100 psi for proper operation Gas Expired Gas expiration date is located on the gas cylinder Wrong gas concentration s Wrong gas type If a monitor fails 3 consecutive t...

Страница 10: ...nload the Bump Calibration Station Manager software from the following link 32TUhttps www macurco com ddownload 2090U32T Save the file to a place you will remember 5 1 2 Install the software To install the software 1 Navigate to the saved file 2 Double click the file to run the installer 3 You may receive a security warning Click Run or Run Anyway to continue with the installation 4 The Docking Ma...

Страница 11: ...on REV 1 0 34 2900 0513 2 10 P a g e 5 Now select the installation folder The default folder is the best option for most people Click Next 6 The installation is ready to begin Click Install 7 You may receive a warning Click Yes to continue ...

Страница 12: ...k or the included USB Disk can be used 5 2 1 Connect Using PM Link The Bump Calibration Station can be conveniently configured using the PM Link device The unit can communicate with the PM Link device using the IR communication in Unit 1 bay To connect using the PM Link 1 Plug the PM Link device into a USB port on your computer 2 Place the PM Link device in Unit 1 bay with the IR lens of the PM Li...

Страница 13: ...B Disk The USB Disk will need to be removed from the Bump Calibration Station and plugged into a USB port on the computer To connect the USB Disk to the Bump Calibration Station Manager software 1 Turn off the power to the Bump Calibration Station by pressing the Bump Test and Calibration buttons simultaneously 2 Remove the two thumbscrews located between the edge of the case and the handle 3 Lift...

Страница 14: ...on until the Write button is pressed see 5 3 1 1 5 3 1 1 Station Info Serial Number Cannot be changed It uniquely identifies the Bump Calibration Station unit Version The firmware version installed on the Bump Calibration Station Auto Power off Determines the idle time before the Bump Calibration Station automatically switches to sleep mode Can be set from 1 3600 seconds Default is 300 seconds Max...

Страница 15: ...If the Calibrate box is unchecked the calibrate function of the Bump Calibration Station will be disabled With the calibrate function disabled pressing the Calibrate button will not calibrate the monitor but it will update configuration or firmware if the option is enabled The power LED will turn orange for 5 seconds then return to green Bump Tests If the Bump Tests box is unchecked the Bump Test ...

Страница 16: ...is verified 5 3 1 3 Bump Calibration Station Firmware Upgrade To upgrade the Bump Calibration Station firmware 1 Click Browse 2 Navigate to the location of the firmware file 3 Click Open 4 The software interface will show the file path of the new firmware in the Station Firmware Upgrade field Click Write to write the new firmware to the Bump Calibration Station 5 3 2 PM100 Tab The Bump Calibration...

Страница 17: ...rform a Read operation when opening If the device is not in place the Read operation may need to be performed once the software is opened to display the configuration of the device User ID The User ID field can be used to add a custom message Calibration Interval Days The Calibration reminder can be configured from n a Off to 365 days Bump Interval Days The Bump Test reminder can be configured fro...

Страница 18: ...e current PMR 100R configuration options saved in the Bump Calibration Station press Read NOTE Pressing Read will cause any unsaved changes on the PM100 tab to be lost 5 3 2 4 PM100 Firmware Upgrade The Bump Calibration Station can be used to upgrade the PMR 100R firmware The Bump Calibration Station can keep a copy of the PMR 100R firmware saved on its USB Disk PMR 100 Rmonitors require a firmwar...

Страница 19: ... monitor used in the Bump Calibration Station A single log file will be created for the Bump Calibration Station NOTE The Logs tab is not available if connected via USB Disk Instead the log files can be viewed and saved directly by navigating to the Logs folder in the USB Disk 5 3 3 1 View Log Files To view a list of the log files click Refresh while connected to the Bump Calibration Station The l...

Страница 20: ...the desired save location If saving many log files it may be advisable to create a folder for the log files 4 Press OK The download will start 5 When the download is complete a Log data save success message will pop up 5 3 3 3 Log Information The log files are saved in a csv file format They can be opened by any spreadsheet software Each time a test is performed a log similar to the image below wi...

Страница 21: ...n only be set when connected using the PM Link 5 3 4 1 Station Time Set NOTE If the Docking Time Auto Sync option is checked in Station Info the option to manually set the Bump Calibration Station time will not be available To manually set the Bump Calibration Station time 1 Update the date field to the desired time Click CURRENT TIME to set the date field to the current computer time 2 Click DATE...

Страница 22: ...LIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR PARTICULAR PURPOSE ARE LIMITED TO A PERIOD OF TWO 2 YEARS FROM THE PURCHASE DATE Macurco shall not be liable for any incidental or consequential damages for breach of this or any other warranty express or implied arising out of or related to the use of said gas detector The manufacturer or its agent s liability shall be limited to replacement or r...