Manual
Instructions for use
BreastBoards & ThoraxSupports
5
02082021V11IFU122500
MacroMedics ®
2.2
MR CONDITIONAL
The following devices are MR conditional:
• 122750 BreastBoard LX™ MRC
• 122150
ThoraxSupport™, MRC
• 122660
ThoraxSupport™short MRC
• 122740 Wrist support for handpoles set/2
• 174590
OmniBoard™ BTM RE MRC
• 175060
OmniBoard™ BTM, S-Type MRC
The devices labelled as MR Conditional can be safely used in a MR system meeting the following condi-
tions:
•
Static magnetic field of 1.5 Tesla and 3 Tesla,
with
•
Maximum spatial field gradient of 2,500 G/cm(25 T/m)
• Maximum force product of 50,000,000 G²/cm (50 T
2
/m)
Due to size and location of the metallic parts, no clinically significant RF heating is expected.
It is the responsibility of the user to ensure that these conditions are met.
2.3
MR SAFE
The following devices are MR safe:
• 122530
WedgeThoraxSupport™, MR
3. Intended Use
Positioning of the patient during radiotherapy and radiodiagnostics, including MR where indicated.