E14
AREA 1: FOR LAST DISPENSE CYCLE
AREA 2: FOR MEASUREMENT UNITS
AREA 3: FOR ACCUMULATED
TOTAL
AREA 4: FOR BATTERY
CAPACITY
Notes: It will flashing when
battery voltage less than 3.3V;
Menu: MOVING THE NUMBERS FOR
ADJUSTING THE CALIBRATION AND
CHOOSING THE UNITS
Reset: CHOOSING THE UNITS AND SETTING
CORRECTION FACTOR
Display and Button Usage
To Activate the Digital Display
1. Press the RESET key pad to clear the meter before starting a new dispense cycle. This is the
best way to activate the meter, because it also clears the quantity of the last dispense cycle.
2. The digital display can also be activated by pressing the MENU key pad or by running fluid
through the meter.
Calibrating Automatically
1. Press Reset key for 1 second and Area 1 shows .000 L;
2. Keep running fluid through the meter until the fluid running out of the certain measuring
cylinder. The display will show certain value as *.**L, press the Menu key for 3 seconds. The
meter will enter instrument calibration mode and Area 1 shows 00*.**L;
3. Press Menu key to move the flashing digital, press Reset key to choose number, enter the
fluid volume and press Menu key again for 3 seconds, then exit the calibration mode. The
calibration settled.
Show Current Correction Factor
1. Press Menu key and Reset key together. The display will show the correction factor, which
can be reset.
Reset The Resettlement Total
1. Press Menu key for 10 seconds, the accumulated total will be reset to be “0”.
To Change the Measurement Units (L / GAL/ PT / QT meter only)
1. Move the flashing display to Area 2 by pressing Menu key, then press Menu key to choose
measurement unit;
2. Press Menu key over 3 seconds to exit the setting mode.
CAUTION
1. To be sure the proper amount of fluid is dispensed, always use the same measurement unit
for a particular fluid. Units should be changed only by authorized employees.
2. To Verify the Accuracy of an Digital Meter.
3. Use a clean, calibrated container. If using a single container, be sure to clean it after each
dispense.
4. Have pump air pressure at the lowest possible setting for dispensing fluid.
5. Put the tip of the nozzle at the bottom of the calibrated container.
6. If the tip of the dispense valve does not reach the bottom of the calibrated container, use a
length of plastic tubing over the tip of the nozzle to ensure liquid enters the container from the
bottom.
7. Trigger the gun slowly so the fluid immediately covers the tip of the dispense valve.