background image

7

NOTES:

Содержание IM50

Страница 1: ...ht Service Centre or Macnaught P L IMPORTANT INFORMATION The Macnaught IM50 Positive displacement flowmeter have incorporated the oval rotor principal into its design This has proven to be a highly re...

Страница 2: ...Remove the 8 x philips head screws 1 from the meter cover 2 Remove the meter cover 2 and o ring 3 3 Remove both rotors 4 4 Clean and inspect all parts Replace any suspect worn or damaged components N...

Страница 3: ...oint 25 change non resetable accumulative total units 26 change non reset accumulative total decimal point 22 change batch total decimal point 27 change K factor 31 change flowrate units 32 change flo...

Страница 4: ...1 1 decimal place 27 K FACTOR kFac 27 normal setting is 112 00 pulses b Press RESET 7 times kFac 27 will appear c Press and hold down RESET for 2 seconds program will appear d To change the K factor...

Страница 5: ...ds program will appear e Press RESET to select sleep setting no sleep sleep 5 f Press and hold down RESET until program disappears to accept setting f Press and hold down RESET until program disappear...

Страница 6: ...AGNOSTICS dIA 7 diagnostic mode b Press the DISP button 5 times to reach dIA 7 c Press RESET dIA 71 will appear d Press and hold down RESET for 2 seconds Program will appear then all the segments shou...

Страница 7: ...7 NOTES...

Страница 8: ...re not manufactured by Macnaught and are supplied by Macnaught under a brand name other than Macnaught the Warranty period is the longer of 12 months from the date of purchase of the relevant product...