Follow the manufacturer’s safety instructions for the use of the cleaning product 

before use to ensure safe use for the operator and the patient. 

Ensure the cloth is damp before the cleaning process. DO NOT use a wet cloth 

over electrical systems. 

 

Application is through a clean damp cloth applied to wipe the device down. Use in the 

following dilutions to ensure an effective clean: 

 

 

Actichlor™ dissolvable chlorine tablets provide a concentration of 1000 ppm of 

available chlorine (0.1%) per 1 tablet 

 

1 tablet (1.7g formed tablet (x1)) will create a virucidal solution, diluted in 1 litre of 

water to provide effective means to clean a “dirty” device. This is also ideal for use 

after an outbreak of the Norovirus/winter vomiting and can be used as a precaution 

against C.Diff. It is effective against viruses, bacteria, spores, yeasts and moulds. 

 

The contact time against the outer components of the device should be for 5 minutes 

to prevent any virucidal infections without a degradation to the functionality of the 

device. 5 minutes is a recommended contact time. The device can withstand a longer 

contact period but the 5 minute recommendation as a minimum must be followed to 

provide an effective cleaning regime. 

Blood spills should be dealt with by an increased concentration of the solution – please refer 

to the instructions on the manufacturers product labelling. 

 

Содержание TX600

Страница 1: ...993553 Pre Sales Information Rev A P a g e 1 of 7 CONTENTS Pre Sales Information Revision A TX600 Fixed Ceiling Hoist ...

Страница 2: ... Sales Information Rev A P a g e 2 of 7 CONTENTS Contents 1 0 Intended Use 3 2 0 Maintenance 3 3 0 General Cleaning 3 1 Disinfecting if necessary 3 4 0 Technical Specification 5 5 0 Shipping Storage Conditions 7 ...

Страница 3: ...check on the unit based on the requirements held within the manual Authorised Engineer 3 0 General Cleaning The exterior of the TX600 can be cleaned using a damp soapy cloth for general cleaning duties Please ensure the cloth is damp and not wet Ensure the exterior of the device is dry after cleaning Dry using a clean dry cloth Handset dry cloth wipe only Lift tape dry cloth wipe only Carry bar da...

Страница 4: ...idal solution diluted in 1 litre of water to provide effective means to clean a dirty device This is also ideal for use after an outbreak of the Norovirus winter vomiting and can be used as a precaution against C Diff It is effective against viruses bacteria spores yeasts and moulds The contact time against the outer components of the device should be for 5 minutes to prevent any virucidal infecti...

Страница 5: ...es of the hoist The direction of travel can only be made within the boundaries of where the hoist is in the track system There are no necessary modifications required for the device to perform its intended use However should the device or the installed system require modification please consult your local Mackworth dealer to arrange a date and time to assess the required changes to the system Fig ...

Страница 6: ...neumatic Sound Level 54 dB A Traverse Speed SWL 1 2 3 4 40 12mm s 98 24mm s 150 00mm s 196 88mm s Traverse Speed No Weight 1 2 3 4 45 48mm s 104 34mm s 158 32mm s 200 00mm s Traverse Stopping Distances SWL 1 2 3 4 6mm 31mm 58mm 84mm Traverse Stopping Distances No Weight 1 2 3 4 6mm 31mm 60mm 89mm Traverse Duty Cycle 25 use 75 rest 90 seconds use 270 seconds res Lifting Speed SWL 29 52mm s Lifting ...

Страница 7: ...king mechanisms on lift tape 2 5N Spring clips on carry bar 8N Manually traversing fully loaded hoist SWL 50N Manually traversing unloaded hoist No weight 10N Table 3 5 0 Shipping and storage conditions 25 C to 5 C 13ºF to 41ºF with any humidity level 5 C to 35 C 41 F to 95 F at a relative humidity up to 90 35 C to 70 C non condensing at a water vapour pressure up to 50hPa The Intended use of the ...

Отзывы: