Follow the manufacturer’s safety instructions for the use of the cleaning product
before use to ensure safe use for the operator and the patient.
Ensure the cloth is damp before the cleaning process. DO NOT use a wet cloth
over electrical systems.
Application is through a clean damp cloth applied to wipe the device down. Use in the
following dilutions to ensure an effective clean:
•
Actichlor™ dissolvable chlorine tablets provide a concentration of 1000 ppm of
available chlorine (0.1%) per 1 tablet
•
1 tablet (1.7g formed tablet (x1)) will create a virucidal solution, diluted in 1 litre of
water to provide effective means to clean a “dirty” device. This is also ideal for use
after an outbreak of the Norovirus/winter vomiting and can be used as a precaution
against C.Diff. It is effective against viruses, bacteria, spores, yeasts and moulds.
•
The contact time against the outer components of the device should be for 5 minutes
to prevent any virucidal infections without a degradation to the functionality of the
device. 5 minutes is a recommended contact time. The device can withstand a longer
contact period but the 5 minute recommendation as a minimum must be followed to
provide an effective cleaning regime.
Blood spills should be dealt with by an increased concentration of the solution – please refer
to the instructions on the manufacturers product labelling.