Service, maintenance, and transport
Warning!
Do not remove the cover of the reader� Injury to eyes from UV
light can occur� All internal service and maintenance work must be done by a
technician approved by Mabtech� Contact Mabtech for more information�
Caution!
Only use shielded signal cables that are shorter than 3 m to
comply with applicable EMC requirements� If the USB cable is longer than 2
m, the connection between the reader and the computer can stop working�
Caution!
Use only the parts specified below. Using other parts will void the
warranty�
Return the equipment in the same box that it arrived in for repair/service/transport�
Environmental conditions for transport and storage should be according to IEC 60068-2
and IEC 60721- 3:
•
Public transportation: -40 to +70ºC
•
Storage locations: -25 to +55ºC
Only the following parts can be replaced by the user:
Part
Description
Power cable
•
240 V/10 A (international) or 100 V/15 A (North America)
•
IEC 60320-1 C13 connector
Ethernet cable
•
Shielded Ethernet cable (SFTP or STP)
•
Cat� 5e or 6A
•
RJ45 male – RJ45 male
•
≤ 3 m
USB cable
•
Shielded USB cable
•
Type A 3�0 male - type B 3�0 male
•
≤ 2m
Fuses x 2
•
Cartridge fuses SP 5×20
•
250V 4A
•
IEC 60127-2
•
Quick-Acting F, H (H = High Breaking Capacity (Ceramic Tube))
Note:
All components in Mabtech IRIS™ are compliant with RoHS(Restriction of
Hazardous Substances Directive, 2011/65/EU)�