NOTE:
Your Nebulizer is equipped with one battery
lamp. It is the first lamp on top of the front.
1) If this lamp,
Fig. 6
, changes from green to
red during use, the rate of nebulization
will decrease, then cease, indicating that the
battery needs to be charged. If this occurs,
(1) Press the POWER
button to turn off.
(2) Replace with a fully
charged battery or
connect to the
adapter.
(3) Press the POWER
button and continue
treatment.
To start the unit,
IMPORTANT!
Your unit has one ‘POWER’ switch
on the front of the unit.
1. Place the unit on a solid, flat dry surface
making sure that the water reservoir and
medicine cup have been properly filled and
the mouth nozzle or other attachments are
properly connected to the cover.
2. Press the ‘POWER’ button.
To adjust the airflow, and nebulizer rate,
1. Your Ultrasonic Nebulizer has 3 airflow
settings: High, Medium and Low. The default
airflow setting is High when turned ON.
2. Based on instructions from
your physician and/or
respiratory therapist,
press the airflow button
to adjust the airflow
volume. The Airflow
Volume Indicator Lights
will show the level (high,
medium or low) selected,
Fig. 7.
To begin a treatment,
1. Follow the instructions provided by your
physician and/or respiratory therapist.
2. Your Nebulizer is designed to start at a
lower atomization rate in the beginning and
will become stronger after a few minutes.
NOTE:
Although many medications are
delivered within six minutes, thicker medications
will take longer to nebulize. The viscosity of the
medication and airflow setting may result in
longer treatment times and reduced battery life
between recharges.
Remember:
• Keep your eyes away from the output of
medication mist.
• If any discomfort or abnormality occurs, stop
using the unit immediately.
• Do not tilt or shake the unit when in operation.
• Opening the cover during operation may cause
a severe injury.
To end a treatment,
1. When the treatment has ended, press the
POWER button. Otherwise, your unit will
automatically shut off after approximately 10
minutes of non-use.
2. Unplug the adapter from the outlet and detach
the adapter plug from the unit’s AC/DC port.
3. Dispose excess medication and water before
disconnecting the battery.
4. Allow the unit to cool — approximately
20-30 minutes.
Remember:
• The unit should not be used for more than
30 minutes at a time.
• Wait 30 minutes before reusing the unit.
After every use,
1. Allow the unit to completely cool.
2. Remove the cover.
3. Carefully remove the medicine cup and
pour out remaining medication.
4. Empty the water reservoir.
5. Gently wipe the water reservoir and
transfer cone with a soft, lint-free cloth.
Remember, the transfer cone should not
be removed.
6. Follow the cleaning procedures provided
in this guidebook.
Fig. 6
Troubleshooting
If any abnormality occurs during use, please check and correct the following:
Condition
Correction
Unit does not
Check the AC adapter connection to the outlet and/or check the AC plug to the
operate when
unit’s AC port.
POWER button
is pressed.
Check that the water reservoir is filled to the correct level.
If using the optional battery pack, check that the battery is completely charged.
Check that the battery is mounted correctly to the unit.
No misting
Check that there is medication in the medicine cup.
or low rate
of misting
Check the medication cup for damage.
Check that the transducer is clean.
Fan is not
Check that the fan is firmly and properly mounted to the motor shaft.
working
properly
The top light
The battery needs to be charged. Nebulization will decrease and then cease.
changed from
green to red
during
nebulization.
Replacement Parts
Replacement parts are available through your local retailer or they can be ordered by mail. See
separate replacement part list for details.
6
7
Fig. 7
POWER
AIR FLOW
High
Medium
Low