Description of Display Symbols
Troubleshooting
If any abnormality occurs during use, please check and correct the following:
Condition
Correction
Display is blank when power is on.
Check and correct the polarity of the
installed batteries. Reinstall or replace
batteries.
Measurement incomplete or
Review and follow “Applying Your
abnormally low or high values
Blood Pressure Cuff” and “Taking
displayed.
Your Blood Pressure Reading” sections.
Replace the batteries.
EE symbol displays.
Do not talk or move during the
measurement. Review “Applying Your
Blood Pressure Cuff” section.
Measurements are different from
Remember, blood pressure readings
those typically measured by
are influenced by physical and mental
physician or every measurement
conditions and/or even the time of day.
reading is different.
Daily measurements should be
interpreted by your physician.
Care and Maintenance
1. Disconnect the cuff tubing from the monitor and remove batteries prior to storing.
2. Only use a soft, dry cloth to clean your blood pressure monitor.
3. Avoid using any liquids on the monitor or cuff.
4. Do not store the unit where it will be exposed to direct sunlight, dust or humidity.
5. Avoid extreme temperatures.
6. Never disassemble the monitor or cuff.
7. Dropping or subjecting your blood pressure monitor to strong shocks should be avoided.
Product Specifications
Name and model number
Ultra Digital Blood Pressure Arm Monitor, 04-596-008
Display system
Digital display/LCD
Measuring method
Oscillometric
Power source
4 “AA” batteries or 6V DC 600mA
Measuring range
Pressure: 0-300 mmHg
Pulse: 30-180 beats/minute
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of reading
Inflation
Pump Driven
Deflation
Automatic Pressure Release Valve
Memory
Built-in memory enabling display of up to 120 (60 per memory
bank) measurements with last three reading average.
Automatic shut-off
Approximately 1 minute after last button operation.
Operation environment
Temperature 41
o
F – 104
o
F (5
o
C – 40
o
C)
Humidity < 85%
Storage environment
Temperature -4
o
F – 131
o
F (-20
o
C – 55
o
C)
Humidity < 95%
Monitor dimensions
6” x 4-1/4” x 2-3/8”
Weight
12 oz. (without batteries)
Arm circumference range
Adult size cuff fits arm circumference: 11-3/4” - 16-1/2” (30-42 cm),
Large Adult Cuff fits arm circumference 16-1/2” - 18-7/8”
(42-48 cm)
Accessories
Adult Cuff, Large Adult Cuff, AC Adapter, Detailed Guidebook,
Quick Start Guide, Blood Pressure Log, Four “AA” Batteries,
Storage case
Options (sold separately)
Small adult size cuff fits arm circumference: 8-5/8” -11-3/4”
(22-30 cm).
Specifications are subject to change without notice.
– 16 –
#91-032-596 10/08
©2008 MABIS Healthcare
– 17 –
Indicates a measurement
within Stage 3 Hypertension.
Indicates a measurement
within Stage 2 Hypertension.
Indicates a measurement
within Stage 1 Hypertension.
Appears when batteries should be
replaced.
Shows the pulse rate per minute.
Occurs when an error was made
during measurement.
Appears when there is an irregular
heartbeat detected during
measurement
Appears when a measurement
error has occurred. Restart the
measurement.
Appears when the cuff is unstable or it
has too much air left in it. Once the
arrow stops flashing, the unit is ready
to take a measurement.
Appears when recalling a
measurement from memory
DIA
SYS
DIA
SYS
DIA
SYS
Indicates the memory user’s memory
bank