M-system PNT Скачать руководство пользователя страница 3

PNT

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201  Fax: +81(6)6659-8510  E-mail: [email protected]

EM-3230  P. 3 / 3

3) Again, with 50% input angle, confirm that output is 50 

±0.5%.  If the output deviates from 50 ±0.5%, repeat the 

steps 1) and 2).

4) Set the input angle to 0%.  

If output deviates from 0%, 

adjust Zero so that output 

becomes 0%. 

  See Figure 3.

5) Set the input angle to 100%.  

If output deviates from 

100%, adjust Zero so as to 

compensate a half of the de-

viation.

  See Figure 4.

6) With 100% input angle, ad-

just Span to compensate the 

other half of the deviation in 

order to obtain 100% output.

  See Figure 5.

7) Apply 0% and 100% input signals and confirm output 

changes accordingly.  If the output deviates from the in-

put, repeat the steps 4) to 6).

Note:  For ease of illustration, a linearity error of the sensor 

angle is expressed in an excessive manner in each graph.

MAINTENANCE

Regular calibration procedure is explained below:

 

CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 25%, 
50%, 75% and 100% input angle.  Check that the output sig-
nal for the respective input angle remains within accuracy 
described in the data sheet.  When the output is out of toler-
ance, recalibrate the unit according to the “ADJUSTMENT 
PROCEDURE” explained earlier.

LIGHTNING SURGE PROTECTION

M-System offers a series of lightning surge protector for 
protection against induced lightning surges.  Please contact 
M-System to choose appropriate models.

OUTPUT %

100

50

0

100
INPUT %

0

Fig. 3

OUTPUT %

100

50

0

100
INPUT %

0

Fig. 4

1/2 deviation

OUTPUT %

100

50

0

100
INPUT %

0

Fig. 5

1/2 deviation

Отзывы: